G
H
I
F
Fiig
gu
urre
e 2
24
4 –– P
Piin
n a
an
nd
d h
hyyd
drra
au
ulliicc ccyylliin
nd
de
err rre
em
mo
ovva
all
13. Remove the hydraulics assembly and properly discard.
N
No
otte
e -- Do not dispose of as unsorted municipal waste. Refer to your local distributor for return or collection systems
available in your country.
14. Reverse steps to reinstall.
15. Verify proper operation before you return the product to service.
H
Hyyd
drra
au
ulliicc ccyylliin
nd
de
err rro
od
d e
en
nd
d rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
1/4'' hex wrench
•
1/2'' combination wrench
•
Diagonal pliers
P
Prro
occe
ed
du
urre
e::
1. Pull the trolley assembly out of the patient compartment until it locks into the loading position.
2. Remove the trolley covers. See
3. Press the main power button to turn the product off.
4. Using a 1/4" hex wrench and a 1/2" combination wrench, remove the end cap cylinder pin (A) and nut (B) (Figure 25).
EN
44
6390-309-002 Rev AB.0
Содержание 6390
Страница 2: ......
Страница 6: ......
Страница 77: ...A An nc ch ho or r a as ss se em mb blly y 6390 001 011 Rev AC Reference only 6390 309 002 Rev AB 0 71 EN ...
Страница 85: ...6390 309 002 Rev AB 0 79 EN ...
Страница 91: ...6390 309 002 Rev AB 0 85 EN ...
Страница 95: ...Torque item B to 68 92 in lb Torque item K to 20 28 ft lb 6390 309 002 Rev AB 0 89 EN ...
Страница 96: ...Torque item DE to 480 600 in lb EN 90 6390 309 002 Rev AB 0 ...
Страница 97: ...Torque item C to 45 max in oz 6390 309 002 Rev AB 0 91 EN ...
Страница 98: ...EN 92 6390 309 002 Rev AB 0 ...
Страница 99: ...6390 309 002 Rev AB 0 93 EN ...
Страница 100: ...EN 94 6390 309 002 Rev AB 0 ...
Страница 101: ...6390 309 002 Rev AB 0 95 EN ...
Страница 102: ...EN 96 6390 309 002 Rev AB 0 ...
Страница 103: ...6390 309 002 Rev AB 0 97 EN ...
Страница 137: ......
Страница 138: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 6390 309 002 Rev AB 0 WCR AE 5 2021 02 ...