If the trolley is unlocking from the transfer:
1. Remove the cot.
2. Move the trolley to the middle of the transfer.
3. On the extended transfer (located just to the head end of the foot end fastener), the transfer trolley pawl (6390-001-073)
should be sticking out of the transfer about 1/2”. If the pawl is not sticking up or does not return when pressed, check its
return spring. Make sure that there is no debris in the return spring.
4. Make sure that the pin of the roller in the pawl is not sticking out on either side.
5. Remove the transfer head end dead stops.
6. Remove the transfer from the anchor.
7. Clear any debris or foreign material from the top of the anchor and the channels in the transfer.
N
No
otte
e -- Foreign objects that do not allow the trolley to reach the full out position could cause the mechanism to not
operate properly.
8. Inspect the trolley stop ramp (6390-001-325) on the underside of the trolley middle mechanism to make sure that it is
present and secure.
If the transfer is unlocking from the anchor:
1. Remove the cot.
2. Move the trolley to the middle of the transfer.
3. Check for debris around the transfer lock trigger (6390-001-231).
4. Push the transfer all the way to the transport position.
5. Pull on the transfer to make sure it is locked.
6. If the transfer is not locked, the transfer lock pin in the anchor may be stuck or broken.
7. Check the return spring and look for burrs or debris.
T
Trro
olllle
eyy w
wiillll n
no
ott u
un
nllo
occkk ffrro
om
m tth
he
e ttrra
an
nssffe
err p
po
ossiittiio
on
n
1. Remove the cot.
2. Using the head end transfer lock override slide (6390-001-261), move the transfer in and put the trolley in the mid
position.
T
Trro
olllle
eyy p
pu
um
mp
p rru
un
nss ffo
orr a
an
n e
exxtte
en
nd
de
ed
d ttiim
me
e a
afftte
err cco
ott jjo
og
g u
up
p
1. Check the angle position sensor (APS) (6390-001-397) calibration set point. Contact Stryker Technical Support.
2. Remove the patient right side cover. Make sure that the hydraulic cylinder is not “over-stroking.”
N
No
otte
e -- A ~1/8" back settle in the rod of the hydraulic cylinder at full extension is a sign of the over-stroke condition
3. Check for damage of components in the chain of load bearing parts from the lifting arms to the patient right side plate
(6390-001-344).
C
Co
ott d
do
oe
ess n
no
ott llo
occkk iin
ntto
o tth
he
e ttrro
olllle
eyy a
an
nd
d L
LE
ED
Dss fflla
assh
h a
am
mb
be
err
1. Remove the cot from P
Po
ow
we
err--L
LO
OA
AD
D.
2. Make sure that the cot is a P
Po
ow
we
err--L
LO
OA
AD
D compatible cot and that the load wheel horns and pins (6500-002-104) are
present, properly tightened, and in good condition.
EN
18
6390-309-002 Rev AB.0
Содержание 6390
Страница 2: ......
Страница 6: ......
Страница 77: ...A An nc ch ho or r a as ss se em mb blly y 6390 001 011 Rev AC Reference only 6390 309 002 Rev AB 0 71 EN ...
Страница 85: ...6390 309 002 Rev AB 0 79 EN ...
Страница 91: ...6390 309 002 Rev AB 0 85 EN ...
Страница 95: ...Torque item B to 68 92 in lb Torque item K to 20 28 ft lb 6390 309 002 Rev AB 0 89 EN ...
Страница 96: ...Torque item DE to 480 600 in lb EN 90 6390 309 002 Rev AB 0 ...
Страница 97: ...Torque item C to 45 max in oz 6390 309 002 Rev AB 0 91 EN ...
Страница 98: ...EN 92 6390 309 002 Rev AB 0 ...
Страница 99: ...6390 309 002 Rev AB 0 93 EN ...
Страница 100: ...EN 94 6390 309 002 Rev AB 0 ...
Страница 101: ...6390 309 002 Rev AB 0 95 EN ...
Страница 102: ...EN 96 6390 309 002 Rev AB 0 ...
Страница 103: ...6390 309 002 Rev AB 0 97 EN ...
Страница 137: ......
Страница 138: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 6390 309 002 Rev AB 0 WCR AE 5 2021 02 ...