A
F
Fiig
gu
urre
e 3
39
9 –– H
Hyyd
drra
au
ulliicc h
ho
osse
e llo
occa
attiio
on
n
4. Using a T25 Torx driver, loosen (do not remove) the two screws (B) that secure the manifold fitting to the left wing plate
N
No
otte
e -- Do not remove the manifold fitting, just loosen the screws to gain better access to the nuts.
A
B
F
Fiig
gu
urre
e 4
40
0 –– M
Ma
an
niiffo
olld
d ffiittttiin
ng
g ssccrre
ew
wss llo
occa
attiio
on
n
5. Using two 9/16" combination wrenches, disconnect the bad hose from the manifold fitting. Discard the hose.
N
No
otte
e -- Hydraulic fluid will leak from the cylinder and hoses. Lay down towels to catch fluid.
6. Reverse steps to reinstall.
7. Check the fluid level before you return the product to service. See
8. Verify proper operation before you return the product to service.
P
Pu
um
mp
p//m
mo
otto
orr a
asssse
em
mb
bllyy rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
(2) 9/16'' combination wrench
6390-309-002 Rev AB.0
55
EN
Содержание 6390
Страница 2: ......
Страница 6: ......
Страница 77: ...A An nc ch ho or r a as ss se em mb blly y 6390 001 011 Rev AC Reference only 6390 309 002 Rev AB 0 71 EN ...
Страница 85: ...6390 309 002 Rev AB 0 79 EN ...
Страница 91: ...6390 309 002 Rev AB 0 85 EN ...
Страница 95: ...Torque item B to 68 92 in lb Torque item K to 20 28 ft lb 6390 309 002 Rev AB 0 89 EN ...
Страница 96: ...Torque item DE to 480 600 in lb EN 90 6390 309 002 Rev AB 0 ...
Страница 97: ...Torque item C to 45 max in oz 6390 309 002 Rev AB 0 91 EN ...
Страница 98: ...EN 92 6390 309 002 Rev AB 0 ...
Страница 99: ...6390 309 002 Rev AB 0 93 EN ...
Страница 100: ...EN 94 6390 309 002 Rev AB 0 ...
Страница 101: ...6390 309 002 Rev AB 0 95 EN ...
Страница 102: ...EN 96 6390 309 002 Rev AB 0 ...
Страница 103: ...6390 309 002 Rev AB 0 97 EN ...
Страница 137: ......
Страница 138: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 6390 309 002 Rev AB 0 WCR AE 5 2021 02 ...