5. Reverse steps to reinstall.
6. Verify proper operation before you return the product to service.
P
Prre
essssu
urre
e cco
om
mp
pe
en
nssa
atte
ed
d ffllo
ow
w cco
on
nttrro
oll vva
allvve
e rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
1/4'' hex wrench
•
Needle nose pliers
P
Prro
occe
ed
du
urre
e::
1. Pull the trolley assembly out of the patient compartment until it locks into the loading position.
N
No
otte
e -- This makes working on the product easier, but is not required.
2. Remove the trolley covers. See
3. Remove the hydraulics assembly. See
Hydraulics assembly removal and replacement
4. Using a 1/4" hex wrench, remove the hex plug (A) from the top of the hydraulic manifold assembly (Figure 44).
N
No
otte
e -- Pay attention to the orientation of the valve for reinstallation.
EN
58
6390-309-002 Rev AB.0
Содержание 6390
Страница 2: ......
Страница 6: ......
Страница 77: ...A An nc ch ho or r a as ss se em mb blly y 6390 001 011 Rev AC Reference only 6390 309 002 Rev AB 0 71 EN ...
Страница 85: ...6390 309 002 Rev AB 0 79 EN ...
Страница 91: ...6390 309 002 Rev AB 0 85 EN ...
Страница 95: ...Torque item B to 68 92 in lb Torque item K to 20 28 ft lb 6390 309 002 Rev AB 0 89 EN ...
Страница 96: ...Torque item DE to 480 600 in lb EN 90 6390 309 002 Rev AB 0 ...
Страница 97: ...Torque item C to 45 max in oz 6390 309 002 Rev AB 0 91 EN ...
Страница 98: ...EN 92 6390 309 002 Rev AB 0 ...
Страница 99: ...6390 309 002 Rev AB 0 93 EN ...
Страница 100: ...EN 94 6390 309 002 Rev AB 0 ...
Страница 101: ...6390 309 002 Rev AB 0 95 EN ...
Страница 102: ...EN 96 6390 309 002 Rev AB 0 ...
Страница 103: ...6390 309 002 Rev AB 0 97 EN ...
Страница 137: ......
Страница 138: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 6390 309 002 Rev AB 0 WCR AE 5 2021 02 ...