EN-3
Before each use, check the outer surface of the
endoscope to ensure there are no rough surfaces, sharp
edges, or protrusions.
1
2
4
3
3. Release the endobody clamp to secure the endoscope.
4. Attach a light cable from the light source to the light post on the
endoscope
3
.
5. With the camera console turned on, rotate the coupler’s focusing
ring
4
to establish a sharp focus on the display monitor.
Note:
to remove the coupler, grip the rear adapter and unscrew the
coupler counterclockwise from the camera head.
Reprocessing
These reprocessing instructions are provided in accordance with
ISO 17664, ISO 15883 AAMI TIR12, and AAMI TIR30. The instructions
have been validated by Stryker as being capable of preparing the
device for re-use. To achieve the desired result, the processor shall
ensure that the following instructions are performed as written
in their entirety and as appropriate in the processor’s facility. This
normally requires routine monitoring and validation of the facility’s
reprocessing procedures. Stryker recommends users observe these
standards when reprocessing medical devices.
Overview
Reprocessing the device involves manual or automated cleaning
with either an enzymatic or a non-enzymatic detergent, optional
disinfection, and sterilization.
• Step 1
(required): Cleaning with Enzymatic or Non-Enzymatic
Detergent
• Step 2
(optional): Disinfection
• Step 3
(required): Sterilization
Warnings
• This device must be cleaned and sterilized prior to the first use and
after every subsequent use.
• Use only the sterilization cycles outlined in this document. Using
unspecified sterilization cycles can damage the device or result in
incomplete sterilization.
• Separate the camera head, coupler, and endoscope prior to
cleaning, disinfection, and sterilization. Failure to follow this
instruction will render the devices non-sterile. (Refer to the camera
head and endoscope product manuals for reprocessing instructions
of those devices.)
Содержание 1688-020-122
Страница 1: ...4K Coupler C Mount with Advanced Imaging Modality REF 1688 020 122...
Страница 2: ......
Страница 4: ......
Страница 16: ......
Страница 150: ...EL 146 1 2 1688 P38139 P38888 3 4 5 6 7 8 9 Stryker...
Страница 151: ...EL 147 1 2 1 2 1 2 4 3 3 4 3 5 4...
Страница 152: ...EL 148 ISO 17664 ISO 15883 AAMI TIR12 AAMI TIR30 Stryker Stryker 1 2 3 Stryker Stryker...
Страница 153: ...EL 149 4K autoclave 1 2 3...
Страница 155: ...EL 151 5 Kimberly Clark KC600 KIMGUARD 6 FDA 7 0233 032 301 0233 032 302 0233 410 002 Stryker 4K 1 2 1 2 3 4 5 6 7...
Страница 156: ...EL 152 8 30 9 10 1 10 1 2 3 4 5 6 7 8 30 9 10 1 10 1...
Страница 157: ...EL 153 2 2 1 5 60 C 1 2 2 115 C 3 4 1 4 2 4 1 2 3 4 5 6...
Страница 159: ...EL 155 Stryker Stryker 18 60 C 15 90 4K 2 Stryker Stryker Stryker...
Страница 160: ...EL 156 Stryker...
Страница 172: ...TR 168...
Страница 174: ...RU 170 1 2 1688 P38139 P38888 3 4 5 6 7 8 9 Stryker...
Страница 175: ...RU 171 1 2 1 2 1 2 4 3 3 4 3 5 4...
Страница 176: ...RU 172 ISO 17664 ISO 15883 AAMI TIR12 AAMI TIR30 Stryker Stryker 1 2 3 Stryker...
Страница 177: ...RU 173 Stryker 4K autoclave...
Страница 180: ...RU 176 1 2 1 2 3 4 5 6 7 8 30 9 10 1 10...
Страница 181: ...RU 177 1 2 3 4 5 6 7 8 30 9 10 1 10 1 2 2 1 5 60 C 1 2 2 115 C...
Страница 182: ...RU 178 3 4 1 4 2 4 1 2 3 4 5 6 STERRAD 1 2 STERRAD...
Страница 184: ...RU 180 Stryker Stryker 18 60 C 15 90...
Страница 185: ...RU 181 4K 2 Stryker Stryker Stryker...
Страница 186: ...RU 182 Stryker...
Страница 188: ...CHS 184 8 9 Stryker 1 2 1 2 1 2 4 3 3 4 3 5 4...
Страница 189: ...CHS 185 ISO 17664 ISO 15883 AAMI TIR12 AAMI TIR30 Stryker Stryker 1 2 3 Stryker Stryker 4K...
Страница 192: ...CHS 188 1 2 3 4 5 6 7 8 30 9 10 1 10 1 2 2 1 5 60 C 1 2 2 115 C 3 4 1 4...
Страница 193: ...CHS 189 2 4 5 6 7 8 9 10 STERRAD 1 2 STERRAD 3 4 STERRAD 100S NX 100NX DUO NX ALLClear 100NX ALLClear DUO 5...
Страница 195: ...CHS 191 4K 2 Stryker Stryker Stryker...
Страница 196: ...CHS 192 Stryker...
Страница 198: ...JA 194 5 6 7 8 9 Stryker 1 2 1 2...
Страница 199: ...JA 195 1 2 4 3 3 4 3 5 4 ISO 17664 ISO 15883 AAMI TIR12 AAMI TIR30 Stryker Stryker 1 2 3 Stryker...
Страница 200: ...JA 196 Stryker 4K autoclave 1 2 3 40psi 275kPa...
Страница 202: ...JA 198 1 2 1 2 3 4 5 6 7 8 30 9 10 1 10...
Страница 203: ...JA 199 1 2 3 4 5 6 7 8 30 9 10 1 10 1 2 2 1 5 60 C 1 2 2 115 C 3 4 1 4...
Страница 204: ...JA 200 2 4 1 2 3 4 5 6 STERRAD 1 2 STERRAD 3 4 STERRAD 100S NX 100NX DUO NX ALLClear 100NX ALLClear DUO 5...
Страница 206: ...JA 202 4K 2 Stryker Stryker...
Страница 207: ...JA 203 Stryker...
Страница 208: ...JA 204...
Страница 210: ...KO 206 4 5 6 7 8 9 Stryker Operating the Endobody Clamp Setting Up the Coupler 1 2 1 2...
Страница 212: ...KO 208 3 Stryker Cautions 4K autoclave Limitations on Reprocessing Materials and Equipment 1 2 3 275kPa 40psi...
Страница 215: ...KO 211 5 6 7 8 30 9 10 1 10 Automated Cleaning 1 2 2 1 5 60 C 1 2 2 115 C 3 4 1 4...
Страница 218: ...KO 214 Service Life 4K 2 Stryker Stryker Stryker Disposal...
Страница 219: ...KO 215 Symbol Stryker Stryker...
Страница 220: ......
Страница 221: ......