36
www.stryker.com
EN
0702-002-700 Rev-S
Specifications
Guidance and manufacturer's declaration - electromagnetic immunity
The Neptune 2 ULTRA Rover REF 0702-001-000 is intended for use in the electromagnetic environment specified below. The customer or the user of the
Neptune 2 ULTRA Rover REF 0702-001-000 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1kV for input/output
lines
± 2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality should be that of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5%
U
T
(>95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
<5%
U
T
(>95% dip in
U
T
)
for 5 s
<5%
U
T
(>95% dip in
U
T
)
for 0.5 cycle
40%
U
T
(60% dip in
U
T
)
for 5 cycles
70%
U
T
(30% dip in
U
T
)
for 25 cycles
<5%
U
T
(>95% dip in
U
T
)
for 5 s
Mains power quality should be that of a typical commercial or hospital
environment. If the user of the Neptune 2 ULTRA Rover REF 0702-001-000
requires continued operation during power mains interruptions, it is recommended
that the Neptune 2 ULTRA Rover REF 0702-001-000 be powered from an
uninterruptible power supply or a battery.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristics of a typical
location in a typical commercial or hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.
0000016914, Rev. S Effective Date: Mar 11, 2016 9:46:10 AM
Printed Date: Feb 24, 2020 18:26:59 GMT