0702-002-700 Rev-S
EN
www.stryker.com
19
CAUTION:
ALWAYS use the correct power cord. Configurations may
vary.
e. Connect the rover to facility electrical power using the appropriate
power cord.
f. Push the power switch to the ON position.
g. Read the WARNING message on the control panel display, then
push the control panel display button next to OK to access the
CONTROL screen (Figure 21).
NOTE:
If the 4-liter or 20-liter suction limit setting dial is positioned
in the high suction range, the rover will provide an audible and visual
indication of this condition. See the
High Suction Indicator
section.
1f
1e
1d
1c
1b
1a
Figure 20 To Connect Power
WARNING
HIGH SUCTION DEVICE
DO NOT CONNECT DIRECTLY
TO CHEST TUBES OR OTHER
CLOSED WOUND DRAINS
OK
1g
Figure 21 High Suction Device Warning
2. To Prepare For Fluid Suction
(Figure 22)
NOTE:
The fluid suction tubing and suction accessory are applied
parts.
a. Install new disposable manifold(s) into the manifold port(s). Make
sure the manifold(s) are fully twisted and locked into place.
See the
instructions for use
supplied with the manifold for more
information.
b. Attach the fluid suction tubing to the port(s) of the installed
manifold(s). ALWAYS close unused manifold ports.
c. Attach a fluid suction accessory to the end of the suction tubing
if required.
d. Push the RESET VOLUME button to reset the fluid display level
value to zero if required.
e. If the RESET VOLUME button is pushed, a confirmation screen
will appear. Push the control panel display button next to OK to
reset the volume levels to zero (Figure 23).
2b
2d
2c
2a
Figure 22 To Prepare for Fluid Suction
PRESSING OK
WILL RESET BOTH
VOLUME LEVELS
TO ZERO
OK
ESC
NOTE
2e
Figure 23 To Reset Volume Level to Zero