To Adjust the Rover Settings (continued)
5. From a SYSTEM SETTING screen, push the buttons next to the
arrow icons to highlight or adjust the appropriate setting option.
See the
System Setup Options
table for available setting options.
6. Push the button next to the OK icon to select the appropriate
system setting option (Figure 19). Push the button next to the
ESC icon to cancel the selection and exit the screen.
mL
L
cc
OK
ESC
VOLUME DISP UNITS
Figure 19 Sample System Setting Screen
NOTE:
To reset the smoke evacuator filter timer or fluid suction filter
timer, see the
instructions for use
supplied with the filter for more
information.
System Setup Options
FLUID LEVEL
DISPLAY
OPTIONS
FACTORY DEFAULT
VOLUME DISPLAY
UNITS
mL [milliliter]
mL
L [liter]
cc [cubic centimeter]
VOLUME DISPLAY
BRIGHTness
0 - 100%
75%
FLUID AUDIBLE
INDICATORS
4L
canister
0 - 2000
mL
600 mL (remaining
capacity)
20L
canister
0 - 10,000
mL
2000 mL (remaining
capacity)
CONTROL PANEL
DISPLAY
OPTIONS
FACTORY DEFAULT
VACUUM DISPLAY
BRIGHTness
0 - 100%
75%
VACUUM DISPLAY
CONTRAST
0 - 100%
70%
MAINTENANCE
OPTIONS
FACTORY DEFAULT
SMOKE FILTER
LIFE
Reset timer to accept
80 hours
SUCTION FILTER
LIFE
Reset timer to accept
500 hours
ERROR HISTORY
See the
Error Code
Messages
section.
Not Applicable
NOTE:
After adjusting the rover system settings, disconnect the rover
from facility power and wrap the power cord around the power cord
bracket. The rover is now ready to be used in an operating room
setting.
Before the Procedure
To Setup the Rover
WARNINGS:
▪
Upon initial receipt and before each use, inspect each component
for damage. DO NOT use any equipment if damage is apparent
or the inspection criteria are not met. See the
Inspection and
Maintenance
section for inspection criteria. DO NOT use the
rover until it has been tested properly to ensure functionality. See
the
To Test the Rover
section.
▪
ALWAYS clean the equipment as indicated upon initial receipt and
before each use. DO NOT place the rover within the sterile field.
Failure to comply may cause infection and result in patient or
healthcare staff injury.
CAUTION:
DO NOT clamp or attach any accessory onto the pole or
base of the powered IV pole assembly.
NOTES:
▪
DO NOT use the rover until the docker has been installed and
tested properly to ensure functionality. See the
instructions for use
supplied with the docker.
▪
The volume of the initial prefill fluid in the canister(s) is
accounted for in the fluid level value indicated on the fluid level
display. Foam does not affect this fluid level value.
▪
ALWAYS close unused manifold ports, and remove or clamp
unused tubing to maintain optimal suction levels. The suction level
in each canister is degraded by the suction demand on the other
canister.
1. To Connect Power
(Figure 20)
a. Position the rover on a flat surface and in a convenient location
within the operating room.
b. Lock the rover’s two rear casters to prevent rover movement.
c. Orient the rover’s fluid level display for optimal viewing.
d. Open the canister access doors from each side of the rover to
allow viewing of the contents.