4. To Test Smoke Evacuation
NOTES:
▪
Make sure a smoke evacuator filter is installed in the rover before
testing smoke evacuation. See the
Accessories
section. See the
instructions for use
supplied with the smoke evacuator filter for
installation information.
▪
The rover tracks the life expectancy of the smoke evacuator
filter. DO NOT relocate the smoke evacuator filter after initial
installation into the rover.
a. Push the EVACUATE SMOKE button (Figure 15).
4a
Figure 15 To Test Smoke Evacuation
b. From the control panel display, push the buttons next to the arrow
icons to adjust the smoke evacuator power (Figure 16). Make
sure the smoke evacuation functions correctly. If not, see the
Troubleshooting
section and the
Error Code Messages
section.
4b
SMOKE
20%
AUTO
Figure 16 To Adjust Smoke Evacuation
NOTE:
See the
To Adjust the Rover Settings
section
to make setting
changes as required.
To Adjust the Rover Settings
1. Make sure the power cord is connected between the rover and
facility power.
2. Make sure the power switch is in the ON position and the HIGH
SUCTION DEVICE WARNING has been acknowledged.
3. From the control panel, push and hold the push button until the
SYSTEM SETUP screen appears (Figure 17).
Figure 17 To Access System Setup
4. From the SYSTEM SETUP screen, push the buttons next to the
arrow icons to highlight the appropriate system setting. Push
the button next to OK to view the desired system setting screen
(Figure 18).
OK
ESC
SYSTEM SETUP
VOLUME DISPLAY UNITS
VOLUME DISPLAY BRIGHT
FLUID AUDIBLE INDICATORS
VACUUM DISPLAY BRIGHT
VACUUM DISPLAY CONTRAST
SMOKE FILTER LIFE
SUCTION FILTER LIFE
ERROR HISTORY
Figure 18 System Setup Screen