0702-002-700 Rev-S
Instructions
Before First Use
To Unpack the Rover
WARNING:
HEAVY EQUIPMENT - ALWAYS have more
than one person unpack and move this equipment from
the shipping pallet. See the
Specifications
section for
rover weight. Failure to comply may result in personal
injury.
1. Remove the exterior packaging materials from the rover and
recycle the material as required.
2. Using at least two people, remove the rover from the shipping
pallet.
3. Inspect the rover and components for damage. If damage is
apparent, DO NOT use the equipment.
COMPONENTS
REF
120 VAC ULTRA Rover
0702-001-000
Domestic Power Cord
6000-115-160
Instructions for Use (IFU)
0702-002-700
Neptune 2 Docking Station
Instruction Poster
9100-001-179
To Initially Dock the Rover
NOTE:
Initial docking of the rover is required to replace the transport
fluid in the canisters with prefill fluid. Failure to comply will result in
prefill errors and prevent the rover from operating properly. See the
To Dock the Rover
section to perform initial docking of the rover.
To Test the Rover
1. To Connect Power
a. Connect the rover to facility electrical power using the power
cord. Insert the plug of the power cord through the strain relief
hole before connecting the cord to the rover (Figure 11).
b. Push the power switch to the ON position.
1a
1b
Figure 11 Power Cord Strain Relief
c. Read the WARNING message on the control panel display, then
push the control panel display button next to OK to access the
CONTROL screen (Figure 12).
WARNING
HIGH SUCTION DEVICE
DO NOT CONNECT DIRECTLY
TO CHEST TUBES OR OTHER
CLOSED WOUND DRAINS
OK
1c
Figure 12 High Suction Device Warning
2. To Test Fluid Suction
a. Push the SUCTION button to start the vacuum pump
(Figure 13). Make sure the vacuum pump starts. If not, see the
Troubleshooting
section and the
Error Code Messages
section.
b. Adjust the SUCTION LIMIT SETTING dial of the canister to
be tested to the maximum suction level. Make sure the value
displayed on the control panel changes and reaches a maximum
suction level between 475 and 480 mm-Hg. If not, see the
Troubleshooting
section and the
Error Code Messages
section.
c. Repeat step b to test the other canister.
2a
2b
2c
Figure 13 To Test Fluid Suction
3. To Test the IV Pole
Push the UP and DOWN ARROW buttons to raise and lower the IV
pole (Figure 14). Make sure the IV pole functions correctly. If not, see
the
Troubleshooting
section and the
Error Code Messages
section.
3
Figure 14 To Test IV Pole
0000016914, Rev. S Effective Date: Mar 11, 2016 9:46:10 AM