0702-002-700 Rev-S
EN
www.stryker.com
33
Error Code Messages
NOTES:
▪
Error code messages appear on the control panel display. From
the control panel display, push the button next to the OK icon to
view the action message. To gain access to error code message
history, see the
To Adjust Rover Settings
section.
▪
For service, contact your Stryker sales representative or call
Stryker Neptune Customer Service. See the
Contact Information
section. Outside the US, contact your nearest Stryker subsidiary.
CODE MESSAGE
ACTION
0.X
SYSTEM ERROR
CALL SERVICE
1.X
MEMORY ERROR
CALL SERVICE
2.X
SYSTEM ERROR
CALL SERVICE
3.0
DOCKER COMM
ERROR
REMOVE ROVER - CYCLE
DOCKER POWER - WAIT 60
SECONDS - DOCK ROVER
AGAIN
3.1
COUPLING ERROR
REMOVE ROVER - CYCLE
DOCKER POWER - WAIT 60
SECONDS - DOCK ROVER
AGAIN
3.2
COUPLING ERROR
CALL SERVICE
3.3
OFFLOAD ERROR
REMOVE MANIFOLDS FROM
CANISTERS - DOCK ROVER
AGAIN
3.4
OFFLOAD ERROR
CALL SERVICE
3.5
PREFILL ERROR
CHECK FACILITY WATER
SUPPLY-DOCK ROVER AGAIN
3.6
PREFILL ERROR
CHECK FACILITY WATER
SUPPLY-DOCK ROVER AGAIN
3.7
COUPLING ERROR
CALL SERVICE
3.9
COUPLING ERROR
CALL SERVICE
3.12
DOCK INCOMPLETE
DOCK ROVER AGAIN
3.13
DOCKER NOT READY
REMOVE ROVER - CYCLE
DOCKER POWER - WAIT 60
SECONDS - DOCK ROVER
AGAIN
4.0
LARGE CAN FULL
SMALL CANISTER VOLUME
WILL NOT FIT IN LARGE -
DOCK ROVER WHEN ABLE
4.1
VACUUM ERROR - SM
CALL SERVICE
4.2
EMPTY TANK ERROR
CALL SERVICE
4.3
EMPTY TANK ERROR
DOCK ROVER TO REMOVE
DEBRIS FROM SMALL
CANISTER
4.4
PREFILL ERROR
PREFILL TANK IS EMPTY -
DOCK ROVER
4.5
PREFILL ERROR
PREFILL TANK IS EMPTY -
DOCK ROVER
5.0
FILTER EXPIRED
REPLACE FLUID SUCTION
HEPA FILTER
5.1
FILTER EXPIRED
REPLACE SMOKE FILTER
6.X
EMPTY TANK ERROR
CALL SERVICE
CODE MESSAGE
ACTION
7.0
LEVEL SENSOR
ERROR
CALL SERVICE
7.1
LEVEL SENSOR
ERROR
CALL SERVICE
7.2
LEVEL SENSOR
ERROR
CALL SERVICE
7.7
PREFILL ERROR
CHECK FACILITY WATER
SUPPLY - DOCK ROVER AGAIN
7.8
PREFILL ERROR
CHECK FACILITY WATER
SUPPLY - DOCK ROVER AGAIN
8.X
LEVEL SENSOR
ERROR
CALL SERVICE
9.0
IV POLE ERROR
CALL SERVICE
9.1
IV POLE ERROR
REDUCE POLE LOAD WEIGHT
9.2
IV POLE ERROR
WAIT 10 MIN - RETRY POLE
10.0
SMOKE EVAC ERROR
CALL SERVICE
10.1
SMOKE EVAC ERROR
CALL SERVICE
10.2
NO SMOKE FILTER
SMOKE FILTER NOT DETECTED
- REPLACE SMOKE FILTER
11.X
VOL DISPLAY ERROR
CALL SERVICE
12.X
LEVEL/TEMP SENSOR
ERROR
CALL SERVICE
13.0
VACUUM ERROR - LG
CALL SERVICE
13.1
VACUUM ERROR - LG
CALL SERVICE
13.2
VACUUM ERROR - LG
CALL SERVICE
13.3
VACUUM ERROR - LG
CALL SERVICE
13.4
VACUUM ERROR - LG
CALL SERVICE
13.5
VACUUM ERROR - LG
CALL SERVICE
13.6
VACUUM ERROR - LG
CANISTER IS PRESSURIZED
EXTERNALLY-TURN ON
VACUUM SYSTEM
13.7
VACUUM ERROR - LG
CALL SERVICE
13.8
VACUUM ERROR - LG
CALL SERVICE
13.10
VACUUM ERROR - LG
CALL SERVICE
13.11
VACUUM ERROR - LG
CALL SERVICE
14.X
VACUUM ERROR - LG
CALL SERVICE
15.1
VACUUM ERROR - SM
CALL SERVICE
15.2
VACUUM ERROR - SM
CALL SERVICE
15.3
VACUUM ERROR - SM
CALL SERVICE
15.4
VACUUM ERROR - SM
CALL SERVICE
15.5
VACUUM ERROR - SM
CALL SERVICE
15.6
VACUUM ERROR - SM
CANISTER IS PRESSURIZED
EXTERNALLY-TURN ON
VACUUM SYSTEM
15.7
VACUUM ERROR - SM
CALL SERVICE
15.8
VACUUM ERROR - SM
CALL SERVICE
15.10
VACUUM ERROR - SM
CALL SERVICE
15.11
VACUUM ERROR - SM
CALL SERVICE
16.X
VACUUM ERROR - SM
CALL SERVICE
0000016914, Rev. S Effective Date: Mar 11, 2016 9:46:10 AM
Printed Date: Feb 24, 2020 18:26:59 GMT