Features
Rover Back View (Figure 6)
Q
R
S
T
X
Y
Z
O
N
M
L
U
V
W
P
Figure 6 Rover Back View
L
Single Vacuum Pump (not shown)
–
Creates suction for both
canisters. If the suction level decreases in either canister, due
to an open port, the suction level in the other canister may
decrease.
M
Fluid Suction HEPA Filter Compartment
–
Allows for the
installation and removal of a disposable fluid suction filter.
This filter (included) provides High Efficiency Particulate
Air (HEPA) filtration of the air from the 20-liter and 4-liter
canisters. See the
Inspection and Maintenance
section.
N
Smoke Evacuator ULPA Filter Compartment –
Allows for
the installation and removal of a disposable smoke evacuator
filter (not included) with an Ultra Low Penetrating Air (ULPA)
efficiency rating.
O
Handle
–
Allows for relocation and positioning of the rover.
P
WARNING Flag
–
Allows for quick access to specific HIGH
SUCTION DEVICE WARNING information.
Q
Control Panel
–
Allows the operation of the rover using dials
and push buttons. The panel also provides visual feedback
through a control panel display. See the
Features
,
Control
Panel
section.
R
Suction Chart and WARNINGS Label –
LOW
MED
OFF
HIGH
480
50
0
80
120
Suction Limit
Setting (mm-Hg)
WARNING
ALWAYS use the minimum suction limit
setting required to achieve the desired
clinical outcome.
ALWAYS follow your institution’s guidelines
for suction limits.
DEATH or serious injury can result from
improper suction levels.
WARNING
NOT FOR USE as a suction source for the
following applications:
•
Organ stabilizer/positioner devices
•
Patient positioner devices
•
Intermittent suction applications
DEATH or serious injury can result.
MAX
S
Quick Reference Cards
–
Allows for quick access to specific
system information, warnings, and troubleshooting information.
T
Manifold Holder
–
Allows for the storage of new, unused
manifolds.
U
Auxiliary Suction Ports (optional)
–
Allow connection to a
hospital wall suction regulator if an alternate suction source
is required. See the
Appendix
,
To Use Auxiliary Suction Ports
(optional)
section.
V
High Suction Hazard WARNING Label –
WARNING
HIGH SUCTION HAZARD
Failure to comply could
result in serious injury or
death.
ALWAYS use hospital wall
suction regulator to control
and monitor suction levels
when using auxiliary
suction ports.
W
Auxiliary Suction Ports Label –
(Connect to hospital wall suction regulator)
AUXILIARY SUCTION PORTS
X
Power Cord Receptacle/Switch
–
Connect facility power to
the receptacle using the rover power cord. Push the toggle
switch to apply or remove facility power. See the
Definitions
section for ON and OFF power symbol definitions.
Y
Power Cord Bracket/Cord
–
Store the power cord on the
bracket. Use the power cord to connect the rover to facility
power.
Z
Specification Label –
0000016914, Rev. S Effective Date: Mar 11, 2016 9:46:10 AM
Printed Date: Feb 24, 2020 18:26:59 GMT