FFT
Windowing
2-11
SR780 Network Signal Analyzer
In general, this window is only useful when looking at transients which do not last the
entire time record.
The Uniform window may also be used with signals which are exactly periodic in the
time record such as a chirp or exact bin sine frequencies.
Hanning
The Hanning window is a commonly used window. However, it has an amplitude
variation of about 1.5 dB for signals which are not at exact bin frequencies and provides
only reasonable selectivity. Its side-lobes are very high and broad for off-bin frequencies.
As a result, the Hanning window can limit the performance of the analyzer when looking
at signals close together in frequency and very different in amplitude.
The Hanning window is most often used in noise measurements since it has the lowest
noise floor.
The Hanning window function is
w
i
N
i
=
−
1 0
2
.
cos
π
for i = 0 .. N-1 and N = number of time record points.
Flattop
The Flattop window has the best amplitude accuracy of any window. The amplitude
variation is only about 0.02 dB for signals between exact frequency bins. However, the
selectivity is worse. Unlike the other windows, the Flattop window has a very wide pass
band and very steep rolloff on either side. Thus, signals appear wide but do not leak
across the whole spectrum.
The Flattop window is the best window to use for accurate amplitude measurements.
The Flattop window function is
w
i
N
i
N
i
N
i
N
i
=
−
⋅
+
⋅
−
⋅
+
⋅
1 0 1 93
2
1 29
4
0 388
6
0 028
8
.
.
cos
.
cos
.
cos
.
cos
π
π
π
π
for i = 0 .. N-1 and N = number of time record points.
BMH
The BMH window combines good selectivity and reasonable accuracy (about 0.8 dB for
signals between exact frequency bins). The BMH window has much lower side-lobes
than the Hanning window and very little broadening of non-bin frequencies.
The BMH window is a good window to use for measurements requiring a large dynamic
range.
The BMH window function is
w
i
N
i
N
i
N
i
=
−
⋅
+
⋅
−
⋅
1 0 1 36109
2
0 39381
4
0 032557
6
.
.
cos
.
cos
.
cos
π
π
π
Содержание SR780
Страница 4: ...ii SR780 Network Signal Analyzer ...
Страница 10: ...viii SR780 Network Signal Analyzer ...
Страница 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Страница 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Страница 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Страница 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Страница 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Страница 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Страница 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Страница 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Страница 232: ......
Страница 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Страница 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Страница 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Страница 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Страница 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Страница 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Страница 478: ......