Exceedance Statistics Commands 5-75
SR780 Network Signal Analyzer
Exceedance Statistics Commands
ESTR (?) {i}
The ESTR command sets (queries) the Exceedance Start Index (for both
displays). The parameter i is a waterfall index [0..].
This command is valid only when the Measurement Group is FFT or Octave.
ESTP (?) {i}
The ESTP command sets (queries) the Exceedance Stop Index (for both
displays). The parameter i is a waterfall index [0..]. Remember, the records are
numbered starting from 0. The last record is the number of stored records minus
one. The exceedance calculation can not proceed if the stop index is greater than
the number of records stored in the waterfall buffer minus one.
This command is valid only when the Measurement Group is FFT or Octave.
EPCT (?) {i}
The EPCT command sets (queries) the Exceedance Centile (for both displays).
The parameter i is a centile from 1 to 99.
This command is valid only when the Measurement Group is FFT or Octave.
EXCE d, i
The EXCE command starts the exceedance centile calculation for display d. The
parameter i selects a Trace [1..5] to store the result.
Display d must be paused or done with waterfall storage on and records stored in
the waterfall buffer. Waterfall display is not required to be on.
The result is stored in a data trace and has the same measurement type as the
waterfall measurements.
This command is valid only when the Measurement Group is FFT or Octave.
Содержание SR780
Страница 4: ...ii SR780 Network Signal Analyzer ...
Страница 10: ...viii SR780 Network Signal Analyzer ...
Страница 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Страница 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Страница 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Страница 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Страница 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Страница 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Страница 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Страница 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Страница 232: ......
Страница 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Страница 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Страница 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Страница 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Страница 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Страница 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Страница 478: ......