5-38 Marker
Commands
SR780 Network Signal Analyzer
MRKY ? d
The MRKY ? command queries the vertical axis data value at the marker position
of display d. The returned value is always the absolute position of the marker,
even when the on-screen marker is relative.
This command is not valid if the Marker of display d is Off.
MRKZ ? d
The MRKZ ? command queries the Z position at the marker position of display
d.
If the display is a 2-D display (Nyquist or Nichols view), MRKZ? queries the
horizontal axis data value at the marker position. The returned value is always the
absolute position of the marker, even when the on-screen marker is relative.
MRKY? queries the position along the vertical axis and MRKX? queries the time
or frequency.
If the display is a waterfall, MRKZ? queries the Z-axis record number of the
marker. Lower record numbers are more recent.
This command is not valid if the Marker of display d is Off.
MKMX d
The MKMX command performs Marker to Max on display d. (Same as [Marker
Max] key). The marker of each display must be moved separately (d=2 is not
allowed).
This command is not valid if the Marker of display d is Off.
MKMN d
The MKMN command performs Marker to Min on display d. (Same as [Marker
Min] key). The marker of each display must be moved separately (d=2 is not
allowed).
This command is not valid if the Marker of display d is Off.
MKCN d
The MKCN command performs Marker to Center on display d. (Same as
[Marker Center] key). The center frequency of the FFT span is set to the marker
frequency on display d. The span is decreased if necessary. The center of each
display must be moved separately (d=2 is not allowed).
This command is only valid when the Measurement Group is FFT and display d
is Live. This command is not valid if the Marker of display d is Off.
Содержание SR780
Страница 4: ...ii SR780 Network Signal Analyzer ...
Страница 10: ...viii SR780 Network Signal Analyzer ...
Страница 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Страница 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Страница 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Страница 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Страница 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Страница 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Страница 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Страница 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Страница 232: ......
Страница 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Страница 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Страница 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Страница 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Страница 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Страница 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Страница 478: ......