Waterfall
Display
1-35
SR780 Network Signal Analyzer
Waterfall Display
This example demonstrates the use of waterfall displays. Waterfalls are available for FFT and Octave
measurements for analog inputs as well as capture playback.
In this example, we will simulate a reverberation measurement measuring the SR780 source. To perform
a real measurement, you would use the source to drive a power amplifier and a microphone to receive the
signal.
1. Press [System]
Display the System menu.
Press
<Preset>
Preset returns the unit to its default settings.
Press [Enter] to confirm Preset.
Preset requires confirmation to prevent accidental
reset. Wait until the self tests are completed.
2. Connect the Source Output to the Channel 1 A
Input.
Setup to analyze the source output.
3. Press [Display Setup]
Select the Display Setup menu.
Press <Measure Group>
Change the Measurement Group.
Select (Octave) with the knob and press [Enter].
Choose the Octave group. Both displays are now
making Octave Analysis measurements.
4. Press [Freq]
Select the Frequency menu.
Press <Lowest Band>
Change the lowest band in the display.
Use the knob to select (50 Hz) and press [Enter].
Select 50 Hz as the lowest band. The lowest band
places a limitation on the minimum integration
time. Raising this lowest band allows shorter
integration times.
5. Press [Source]
Select the Source menu.
Press
<Noise>
Choose Noise as the source type. Octave
measurements are generally used to measure noise.
Press
<Type>
Change the type of noise.
Use the knob to select (Pink) and press [Enter].
Choose Pink noise. Pink noise rolls off at -3dB per
octave. This maintains equal power per octave band
and yields a flat octave spectrum.
Press
<Burst>
Change the Burst percentage.
Содержание SR780
Страница 4: ...ii SR780 Network Signal Analyzer ...
Страница 10: ...viii SR780 Network Signal Analyzer ...
Страница 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Страница 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Страница 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Страница 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Страница 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Страница 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Страница 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Страница 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Страница 232: ......
Страница 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Страница 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Страница 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Страница 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Страница 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Страница 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Страница 478: ......