Linking (Advanced Operation) 1-15
SR780 Network Signal Analyzer
Linking (Advanced Operation)
This example investigates the test filter (enclosed with this manual) using FFT measurements. You will
use the SR780 source to provide a broad band source and both displays to measure the output of the
device under test. Display parameter linking and function linking will be explored in greater detail.
1. Press [System]
Display the System menu.
Press
<Preset>
Preset returns the unit to its default settings.
Press [Enter] to confirm Preset.
Preset requires confirmation to prevent accidental
reset. Wait until the self tests are completed.
2. Connect the Source Output to the filter input.
Connect the filter output to the Ch2 A Input.
In this example, only the filter output on Channel 2
is required.
3. Press [Source]
Select the Source menu.
Press
<Noise>
Choose Noise output. The output is random noise.
Press
<Type>
Adjust the Noise Type.
Select (White) with the knob and press [Enter].
This source is White Noise which extends over the
entire 0-102.4 kHz frequency range.
4. Press [Auto Range Ch2]
Let the analyzer automatically set the Input Range
to agree with the signal. Note that the Input Range
readouts at the top of the screen are displayed in
inverse when Auto Range is on.
5. Press [Display Setup]
Select the Display Setup menu.
Press
<Measurement>
Change the Measurement of the active display (A).
Select (FFT ch2) with the knob and press
[Enter].
Choose FFT spectrum of Ch2 for the measurement
in DisplayA (top). Both displays are independently
measuring the filter output spectrum.
6. Press [Freq]
Select the Frequency menu. The menu shows the
frequency parameters for the measurement in
DisplayA (active display).
Содержание SR780
Страница 4: ...ii SR780 Network Signal Analyzer ...
Страница 10: ...viii SR780 Network Signal Analyzer ...
Страница 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Страница 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Страница 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Страница 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Страница 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Страница 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Страница 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Страница 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Страница 232: ......
Страница 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Страница 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Страница 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Страница 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Страница 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Страница 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Страница 478: ......