Frequency
Commands
5-29
SR780 Network Signal Analyzer
Frequency Commands (Swept Sine)
SSTR (?) d {, f}
The SSTR command sets (queries) the swept sine Start frequency of display d.
The parameter f is a frequency (real number of Hz). The range of f is 0.001 to
102.4E3. The set command requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The set
command requires a display to be Live.
SSTP (?) d {, f}
The SSTP command sets (queries) the swept sine Stop frequency of display d.
The parameter f is a frequency (real number of Hz). The range of f is 0.001 to
102.4E3. The set command requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The set
command requires a display to be Live.
SSFR ?
The SSFR command queries which sweep point was the most recently measured.
While a sweep is in progress, SSFR? returns values from 0 to the Number Of
Points - 1.
After a sweep is started, SSFR? will not return 0 until the first point is measured.
This may take some time if Source Ramping is On or if the Settling Time is very
long.
This command is valid only when the Measurement Group is Swept Sine.
SRPT (?) d {, i}
The SRPT command sets (queries) the swept sine Repeat Mode of display d. The
parameter i selects Single Shot (0) or Continuous (1). The set command requires
d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The set
command requires a display to be Live.
SSTY (?) d {, i}
The SSTY command sets (queries) the swept sine Sweep Type of display d. The
parameter i selects Linear (0) or Logarithmic (1). The set command requires d=2
(both displays).
This command is valid only when the Measurement Group is Swept Sine. The set
command requires a display to be Live.
SARS (?) d {, i}
The SARS command sets (queries) the swept sine Auto Resolution Mode of
display d. The parameter i selects Off (0) or On (1). The set command requires
d=2 (both displays).
Содержание SR780
Страница 4: ...ii SR780 Network Signal Analyzer ...
Страница 10: ...viii SR780 Network Signal Analyzer ...
Страница 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Страница 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Страница 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Страница 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Страница 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Страница 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Страница 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Страница 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Страница 232: ......
Страница 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Страница 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Страница 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Страница 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Страница 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Страница 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Страница 478: ......