Remote Programming
mands received at once, queues cleared)
1
Unused
2
Query Error
Set on output queue overflow (too many re-
sponses waiting to be transmitted, queues
cleared)
3
Unused
4
Execution Error
Set when a command can not execute correctly
or a parameter is out of range
5
Command Error
Set when an illegal command is received
6
URQ
Set by any key press or knob rotation
7
PON
Set by power-on
The Standard Event status byte is defined by IEEE-488.2 (1987) and is used primarily to report errors in com-
mands received over the communications interfaces. The bits in this register remain set until cleared by read-
ing them or by the *CLS command.
MCS STATUS BYTE
bit
name
usage
0
Triggered
Set when a record is triggered
1
Plot Complete
Set when a plot is completed
2
Print Complete
Set when a print screen operation is completed
3
Toggled
Set when the accumulate mode toggles between
add and subtract
4
Scan Paused
Set when the unit enters the PAUSE state
5
Inhibited
Set when a record accumulation is inhibited via
the rear panel Inhibit input.
6
Unused
7
Unused
The MCS Status bits stay set until cleared by reading or by the *CLS command.
Содержание SR430
Страница 2: ......
Страница 6: ...4 ...
Страница 22: ...20 Guide To Operation ...
Страница 26: ...24 Guide To Operation ...
Страница 36: ...34 Mode Menu ...
Страница 54: ...52 Save Menu ...
Страница 60: ...58 Recall Menu ...
Страница 70: ...68 Plot Menu ...
Страница 74: ...72 Test Menu ...
Страница 76: ...74 Info Menu ...
Страница 97: ...96 Remote Programming ...
Страница 99: ...98 98 Program Examples ...
Страница 107: ...106 106 Test and Calibration ...
Страница 113: ...112 112 Using Photomultiplier Tubes ...
Страница 125: ...124 124 124 Circuit Description ...