88
Remote Programming
record and i must be greater than the left limit set by LLIM.
OPER (?) {i}
The OPER command sets or reads the type of math operation selected.
The parameter i selects the operation.
i
operation
0
+
1
-
2
x
3
/
4
log
5
sqrt
CALC
The CALC command starts the calculation selected by the OPER com-
mand. This may take some time. Use status byte query to detect when
the calculation is done. Make sure that ARGM or FNAM have been used
to set the argument value before using the CALC command. CALC will
convert the count data into floating point data.
ARGT (?) {i}
The ARGT command sets or reads the argument type. The parameter i
selects Constant (i=0) or File (i=1).
ARGM (?) {x}
The ARGM command sets or reads the constant argument. The parame-
ter x is a real number.
FNAM (?) {string}
The FNAM command sets or reads the active file name for file argu-
ments. All file operations use the name specified by the FNAM com-
mand.
Be sure to use the FNAM string command before any file op-
eration commands.
For example, "FNAM MYDATA.DAT" will set the
active file name to MYDATA.DAT. DOS file name conventions must be
followed, i.e. file names are 8 characters or less with an optional exten-
sion of up to 3 characters. Subdirectories are not supported. All file ac-
cess is to the root directory.
STAT
The STAT command starts the statistical calculations. Only the data with-
in the region specified by the LLIM and RLIM limits is analyzed. Use stat-
us byte query to detect when the statistical calculation is done. If the
STAT command is used before the LLIM and RLIM commands have set
the limits, the results are invalid and an execution error may result.
SPAR ? i
The SPAR? i command queries the results of a statistical calculation. If
no STAT command has been done, the SPAR? command returns invalid
data. The STAT command calculates the mean, standard deviation, total
area, and the baseline area for the data within the limits set by LLIM and
RLIM (see the Math menu section of this manual). The parameter i in the
SPAR? i command selects the parameter to read. The value of i from 0
to 3 selects mean, standard deviation, total area, and baseline area.
Содержание SR430
Страница 2: ......
Страница 6: ...4 ...
Страница 22: ...20 Guide To Operation ...
Страница 26: ...24 Guide To Operation ...
Страница 36: ...34 Mode Menu ...
Страница 54: ...52 Save Menu ...
Страница 60: ...58 Recall Menu ...
Страница 70: ...68 Plot Menu ...
Страница 74: ...72 Test Menu ...
Страница 76: ...74 Info Menu ...
Страница 97: ...96 Remote Programming ...
Страница 99: ...98 98 Program Examples ...
Страница 107: ...106 106 Test and Calibration ...
Страница 113: ...112 112 Using Photomultiplier Tubes ...
Страница 125: ...124 124 124 Circuit Description ...