21
Guide To Operation
KEYPAD
START
STOP
CURSOR
AUTO
RESET
SCALE
7
8
9
PEAK
4
5
6
CENTER
1
2
3
NORM
0
•
--
EXP
ALT
CLR
ENTER
A
B
C
D
G
M
S
H
I
J
N
O
P
T
U
V
Y
Z
+
-
/
*
HELP
LEVELS
MODE
DISPLAY
SETUP
SAVE
MATH
RECALL
TEST
INFO
E
F
K
Q
W
L
R
X
\
(
)
PLOT
=
ENTRY
CONTROL
MENU
SYSTEM
NORMAL AND ALTERNATE KEYS
The normal key definitions are printed on each
key. In addition, each key also has an alternate
definition printed below it. The [ALT] key toggles
the keypad between the two definitions. The ALT
screen indicator is on when the alternate defini-
tions are in use. The [ALT] and [ENTER] keys
have the same definition in both modes. The alter-
nate keys should only be used when accessing
files on the disk drive.
START, STOP, AND RESET
The [START] and [STOP/RESET] keys are used
to start, pause, and reset the counter and accumu-
lation memory.
Data acquisition can be started only if the counters
and memory are in the CLEAR or reset condition;
all data values are zero. The CLEAR condition is
indicated by the CLR data acquisition status indi-
cator. Triggers which arrive while in the CLEAR
state are ignored and SYNC/BUSY and BIN CLK
OUT are inactive.
[START] begins the data acquisition. The first trig-
ger after [START] is pressed will trigger the first
data record. The data acquisition status will switch
to BUSY. Successive triggers will acquire records
which are accumulated. While the unit is BUSY,
the SYNC/BUSY and BIN CLK outputs are active
and accumulated data is graphed on the display.
Each successful trigger will generate a SYNC/
BUSY pulse and a BIN CLK pulse train. Triggers
received during SYNC/BUSY active times are ig-
nored. After the programmed number of records
have been accumulated, data acquisition is halted.
The unit is now in the DONE data acquisition
state. The final accumulated data is now graphed
on the display. Further triggers will be ignored and
SYNC/BUSY and BIN CLK OUT are inactive.
Pressing the [STOP] key while in the BUSY state
pauses the data acquisition. The data acquisition
status will be PAUSE and no triggers will be recog-
nized. Pressing [START] will continue the data ac-
Содержание SR430
Страница 2: ......
Страница 6: ...4 ...
Страница 22: ...20 Guide To Operation ...
Страница 26: ...24 Guide To Operation ...
Страница 36: ...34 Mode Menu ...
Страница 54: ...52 Save Menu ...
Страница 60: ...58 Recall Menu ...
Страница 70: ...68 Plot Menu ...
Страница 74: ...72 Test Menu ...
Страница 76: ...74 Info Menu ...
Страница 97: ...96 Remote Programming ...
Страница 99: ...98 98 Program Examples ...
Страница 107: ...106 106 Test and Calibration ...
Страница 113: ...112 112 Using Photomultiplier Tubes ...
Страница 125: ...124 124 124 Circuit Description ...