10
SR430 Multichannel Scaler
string.
SVTR
Save the Trace Data to the file specified by FNAM.
SVST
Save the Settings to the file specified by FNAM.
SVSS
Save the System Configuration to the file specified by FNAM.
RCTR
Recall the Trace Data from the file specified by FNAM.
RCST
Recall the Settings from the file specified by FNAM.
RCSS
Recall the System Configuration from the file specified by FNAM.
MATH
SMTH i
Smooth the data using 5 pts (0), 11 pts (1), 17 pts (2), 21 pts (3), or 25 pts (4).
FTYP(?) {i}
Set the Fit Type to Line (0), Exponential (1), or Gaussian (2).
FITS
Start the Fit Calculation.
PARS? i
Query Fit Parameter t0 (0), a (1), b (2), or c (3).
LLIM(?) {i}
Set the Left Limit of the curve fit region to bin #i where 0 ≤ i < Right Limit.
RLIM(?) {i}
Set the Right Limit of the curve fit region to bin #i where Left Limit < i ≤ Last bin in record.
CALC
Start a Calculation
ARGT(?) {i}
Set the calculation Argument type to Constant (0), or File (1).
ARGM(?) {i}
Set the Constant Argument to x.
STAT
Start the Statistical Calculations.
SPAR? i
Query Statistical Parameter mean (0), deviation (1), area (2), baseline area (3).
PRINT AND PLOT
PRSC
Print the screen. Same as [PRINT] key.
PALL
Plot the entire screen.
PTRC
Plot the trace only.
PCUR
Plot the cursor only.
DATA ACQUISITION CONTROL
SSCN
Start data acquisition. Same as [START] key.
PAUS
Pause a scan in progress. Same as [STOP] key while BUSY.
CLRS
Reset the data memory. Same as [STOP] key while DONE.
SCAN?
Query the number of records which have been accumulated so far.
DATA TRANSFER
BINA? {i}
Read the entire data record in ASCII format. If i included, read only bin #i data.
BINB?
Read the entire data record in binary format.
INTERFACE
*RST
Reset the unit to its default configurations.
*IDN?
Read the SR430 device identification string.
LOCL(?) {i}
Set the Local/Remote state to LOCAL (0), REMOTE (1), or LOCAL LOCKOUT (2).
STATUS
*CLS
Clear all status registers.
*ESE(?) {i}
Set the Standard Status Byte Enable Register to the decimal value i (0-255).
*ESR? {i}
Query the value of the Standard Status Byte. If i is included, only bit i is queried.
*SRE(?) {i}
Set the Serial Poll Enable Register to the decimal value i (0-255).
*STB? {i}
Query the value of the Serial Poll Status Byte. If i is included, only bit i is queried.
*PSC(?) {i}
Set the value of the Power-On Status Clear bit to Set (1) or Clear (0).
ERRE(?) {i}
Set the Error Status Enable Register to the decimal value i (0-255).
ERRS? {i}
Query the value of the Error Status Byte. If i is included, only bit i is queried.
MCSE(?) {i}
Set the MCS Status Enable Register to the decimal value i (0-255).
MCSS? {i}
Query the value of the MCS Status Byte. If i is included, only bit i is queried.
Содержание SR430
Страница 2: ......
Страница 6: ...4 ...
Страница 22: ...20 Guide To Operation ...
Страница 26: ...24 Guide To Operation ...
Страница 36: ...34 Mode Menu ...
Страница 54: ...52 Save Menu ...
Страница 60: ...58 Recall Menu ...
Страница 70: ...68 Plot Menu ...
Страница 74: ...72 Test Menu ...
Страница 76: ...74 Info Menu ...
Страница 97: ...96 Remote Programming ...
Страница 99: ...98 98 Program Examples ...
Страница 107: ...106 106 Test and Calibration ...
Страница 113: ...112 112 Using Photomultiplier Tubes ...
Страница 125: ...124 124 124 Circuit Description ...