MODE MENU
29
20.48
µ
s
1000
Internal
External
1k
0
More:
Bin Clk Src
Bins/Record
Records/Scan
Trigger Offset
0.00
µ
sec
Bin Width
Toggle Count
Add Toggle Ext
Trigger Offset
Internal Bin Width
Internal External
MODE
Bin Clock Source
Bins/Record
Records/Scan
More
From Offset
Accumulate Mode
Pause at Toggle
ReAcquire
Return
Introduction
The Mode menu sets up the data acquisition parameters, including bin
width, accumulation mode, bins per record, and records per scan. With
the exception of the Records/Scan and the ReAcquire Offset, these pa-
rameters may not be changed unless the unit is in the CLEAR state. Re-
Acquire may only be performed when data acquisition is done.
Bin Clock Source
This key toggles the time base of the time bins from internal clock to ex-
ternal bin clock. When set to External, the EXT BIN CLK input deter-
mines the width of each bin.