SERVO
PRO NANOCHROME Gas Analyser
119
04416001A / Revision 7
12 ANALYSER TROUBLESHOOTING
This section addresses some of the typical issues that may be experienced whilst
operating the SERVOPRO NanoChrome. The section describes the observed
symptoms, the likely causes, and the potential solutions.
If these steps do not resolve the issue, then contact Servomex for technical support.
Symptom #
Symptom description
Cause
1
Baseline noise
1,2,3,4,5,21,24
2
Unstable baseline
1,2,3,4,5,21,24
3
No peak
6,7,8,9,19,22,23,24,26
4
Saturated detector signal
1,2,3,4,19,21,24
5
Negative peaks
1,2,3,4,24
6
NMHC baseline drift
10
7
Poor repeatability on span calibration
1,2,3,4,5,11,12,13,20,21,23,24,25
8
Poor repeatability on process gas
1,2,3,4,5,12,21,23,24,25
9
Results drift on span calibration gas
14,15,23,24
10
Peak elution time moving
11,22,23,24
11
Sensitivity change on span calibration gas
1,2,3,4,16,17,21,23,24,29
12
Span calibration peak not inside window
9,23,24,25,26
13
Unexpected peak on zero gas (blank)
27
14
High carrier flow consumption
28
Table 31 : Troubleshooting Typical Issues
–