SERVO
PRO NANOCHROME Gas Analyser
56
04416001A / Revision 7
2)
Relay Diagnostic
This combo box is used to confirm the Range relays on the I/O Boards are
working properly. Select the I/O Board to diagnose by using the combo box
just under the title “Relay Diagnostic”.
Use the combo box in the state
column of the grid to switch the relay ON or OFF depending on the
configuration in
CONFIGURATION-ADVANCED CONFIGURATION
menu. To apply a change, press
ENTER
on the keypad or perform a mouse
click outside the combo box.
3)
Valves Diagnostic
This combo box is used to confirm the valves on the I/O Boards are working
properly. First select the I/O Board to diagnose by using the combo box just
under the title “Valves Diagnostic”.
Use the combo box in the state column
of the grid to switch the valve ON or OFF. The value is applied after an
ENTER
on the keypad or when a mouse click is performed outside the
combo box.
4)
Status, Alarm 1 and Alarm 2
These group boxes let you change the state of the corresponding relays of
Alarm and Status for a short period of time. If the analyser processes a new
alarm or error, these relays will return to their normal state after a short
moment even if you have triggered a different state. By default, these relays
are on the first I/O board.