2. Preparation: Mounting a Lens and Adjusting the Flange Focal Length
22
When using 1/3 arc shims
Attach shims so that they have the same
thickness in all three locations.
Shims
When using circular shims
Attach with the surface that has markings
on it facing the front.
Markings
Markings
4
Reattach the PL connectors (two
locations) to their original positions, and
tighten the four Phillips screws with 0.18
N·m tightening torque.
5
Reattach the PL-mount flange in its
original position, and tighten the six Torx
screws to a tightening torque of 0.35 N·m
using a T8 torque wrench.
Cleaning the Filter
To clean the filter, first remove the adaptor.
Exercise care when wiping the adaptor
center part (shaded part) with a cloth or
other material to prevent fibers adhering to
surfaces. If fibers are adhering, wipe off using
a soft brush.
Adaptor
Removing the PL Lens Adaptor
Remove the PL lens adaptor when you
want to mount an E-mount lens to the unit.
Attachment/removal is performed with the
rear side of the unit facing down.
Loosen the six hex screws (2.5 mm) and
remove the PL lens adaptor.
PL lens
adaptor
[Notes]
• Removing the battery and placing the rear side of the
unit face down provides stability.
• When attaching/removing the adaptor, take care not
to touch the connector contacts of the unit and PL
lens adaptor.
To attach the PL lens adaptor
1
Check that the mount lever is in the
locked position.
Mount
lever
2
Place the PL lens adaptor back in its
original position, insert the six hex screws
(2.5 mm), and tighten the screws to a
tightening torque of 0.8 ± 0.12 N·m using
a hex wrench (2.5 mm).