OPERATING MANUAL PAT-806 version 1.08
58
Measurement of insulation resistance using test voltage of 100 V
Test range according to IEC 61557-2 for U
N
= 100 V: 100
kΩ…99.9 MΩ
Display range
Resolution
Measurement uncertainty
(basic)
0 k
Ω...1999 kΩ
1
kΩ
(5 % m.v. + 8 digits)
2.00 M
Ω ... 19.99 MΩ
0.01
MΩ
20.0 M
Ω ... 99.9 MΩ
0.1
MΩ
Influencing factor
Designation
Additional uncertainty
Position
E
1
0 %
Supply voltage
E
2
0 %
Temperature
E
3
0.1
%/ºC
Capacity
E
7
0
% for R ≤ 20 MΩ
unspecified for R> 20 M
Ω
Accuracy of generated voltage (Robc [
]
1000*UN [V]): -0 % +30 % from the set value
nominal current: min 1 mA ... 1.4 mA
adjustable lower limit within the range of 0.1
MΩ ….9.9 MΩ with resolution of 0.1 MΩ
adjustable measuring time: "Cont" mode (continuous measurement) 4
s…3 min with a resolu-
tion of 1 s
detection of a dangerous voltage before commencing a measurement
discharging the object tested
Note: For R <50 k
Ω, the uncertainty is not specified.
Measurement of insulation resistance using test voltage of 250 V
Test range according to IEC 61557-2 for U
N
= 250 V: 250
kΩ…199.9 MΩ
Display range
Resolution
Measurement uncertainty
(basic)
0
kΩ...1999 kΩ
1
kΩ
(5 % m.v. + 8 digits)
2.00
MΩ ... 19.99 MΩ
0.01
MΩ
20.0
MΩ...199.9 MΩ
0.1
MΩ
Influencing factor
Designation
Additional uncertainty
Position
E
1
0 %
Supply voltage
E
2
0 %
Temperature
E
3
0.1
%/ºC
Capacity
E
7
0
% for R ≤ 20 MΩ
unspecified for R> 20 M
Ω
Accuracy of generated voltage (Robc [
]
1000*UN [V]): -0 % +30 % from the set value
nominal current: min 1 mA ... 1.4 mA
adjustable lower limit within the range of 0.1
MΩ ….9.9 MΩ with resolution of 0.1 MΩ
adjustable measuring time: "Cont" mode (continuous measurement) 4
s…3 min with a resolu-
tion of 1 s
detection of a dangerous voltage before commencing a measurement
discharging the object tested
Note:
For R <50 k
Ω, the uncertainty is not specified.