OPERATING MANUAL PAT-806 version 1.08
42
Correct result: I
P
< LIMIT
Incorrect result: I
P
> LIMIT
Settings and measurements as in section 4.5.1.
4.11 Rated voltage measurement on welding equipment without
load
Press
U
0
button once or twice depending on
chosen voltage measurement: r.m.s. Ur volt-
age or UP peak voltage. The screen informing
about the readiness for measurement will be
displayed with symbol
or
.
When the limit value must be changed press
SET
.