During the measurement
current I is displayed alter-
natively with the time of
measurement.
Press
ENTER
, if this test is successful (
is displayed) or
STOP/ESC
, when the test result is negative (
is dis-
played).
Notes:
During the measurement in the test socket the voltage of mains is present.
4.14 IEC lead test
Keep pressing
AUTO/IEC
button, until the
screen informing about the readiness for test-
ing IEC conductor is displayed.
When parameters must be changed press
SET
. Setting is performed as in section 4.2.
The following must be successively set:
LIMIT for R
PE
, measurement time R
PE
, LIMIT
for R
ISO
, measurement time R
ISO
.
Connect the mains plug into the test socket and the other into
the IEC socket.