OPERATING MANUAL PAT-806 version 1.08
26
The method of measurement and observations are the same as for the measurement of 200mA.
4.4.2 Compensation of the test lead resistance during the measurement of
the protective lead resistance using 10 A or 25 A current (auto-zero)
In
R
PE
10A
or
25A
mode, press
. Auto-zero
screen will appear with blinking symbol
.
Use the probe connected to
I2
socket and touch PE pin of
the test socket.
Press
START
, to begin auto-zero.
After finishing auto-zero process, for 1s the
message is displayed and the meter en-
ters the measuring function.