Option
Description
Interval (default)
Measurements are initiated in a specified interval.
TRIG input
Measurements are triggered by the positive edge of a DC-voltage signal
applied to the TRIG input (low: 0 ... 0.6 V, high: 2.2 ... 28 V, min. pulse width:
30 ms)
SDI-12/RS-485
Measurements are externally triggered by commands via RS-485 or SDI-12
from ,e.g. a data logger.
all allowed
Measurement is triggered by all options mentioned above.
A measurement interval between 20 sec and 3 hours can be selected.
The type of the serial output protocol. The following options are available:
Option
Description
Sommer (default)
Sommer protocol; data values are returned with an index starting at 1
Standard
Standard protocol; data values are returned without an index
MODBUS
Modbus protocol
The main measurement values are always included in the data output string. Additionally, special
and analysis values can be included. See
for details.
10.4.2 Calibration of SPA-2-ribbon-sensors
The SPA-ribbon-sensors are calibrated by measuring the complex capacities in dry air (zero) and
with an attached reference plate that simulates a certain water content (span). Sommer Mess-
technik only ships calibrated SPA-ribbon sensors and generally, a regular zero-check of the sensors
is sufficient. If required, the reference plate can be obtained from Sommer Messtechnik.
10
C
o
n
fig
u
ra
ti
o
n
Manual
55