The slope applied to the selected variable. Only available if
is set to 1…5 or none. If
is set as S (as source), no scaling is applied.
Offset
The offset applied to the selected variable.
Adjustment (Correction)
A measurement of the selected variable is triggered and the result displayed in the terminal win-
dow. If the measured value deviates from the correct value, the correct value can be entered. This
adjusts the value in
. The factor in
is not affected by this correction.
Test
A measurement of the selected variable is triggered and the result displayed in the terminal win-
dow.
S-TYP
One of the following sensor (or source) types:
S-TYP
Description and S-MEA options
off
No source (deactivated)
SPA-S
Variables of one of the SPA-ribbon sensors (straps)
SWE
Snow water equivalent
Density
Snow density
Ice
Fraction of ice (by volume)
Water
Fraction of water (by volume)
S-Temp
Sensor temperature
C-LF
Capacity at low frequency
C-HF
Capacity at high frequency
Ph-LF
Phase at low frequency
Ph-HF
Phase at high frequency
AIN
Analog input
AN1
Analog input 1 (preconfigured for snow depth measurement)
AN2
Analog input 2
AN3
Analog input 3
AN4
Analog input 4
12
P
ar
am
et
er
d
ef
in
it
io
n
s
Manual
93