Wire color
Pin
Function
Description
white
1
GND
Ground
brown
2
Vsupply
9…27 VDC
green
3
Trigger
low: 0…0.6 V, high: 2…27 V
red
4
IOUT1
Level (4…20 mA)
yellow
-
RS-485 A
RS-485 (1200…115200 Baud)
grey
-
RS-485 B
rosa
-
SDI-12
SDI-12 (1200 Baud)
blau
-
IOUT2
Temperature (4…20 mA)
Table 3:
Connection wires of USH-9
7.3.8
Power supply
The SPA-2 is designed for extreme environmental conditions at remote sites and with no grid con-
nection. The sensor switches automatically into standby-mode between measurements and thus
consumes only approx. 0.2 Ah per day (measurement duration 6 sec and measurement interval 60
sec) which can be supplied by a 12V-solar-generator mounted to the mast.
7.3.9
Signal cables
Please consider the maximum cable lengths for the applied transmission protocol:
Protocol
Max. cable length [m]
SDI-12
60
RS-485
300
Table 4:
Maximum cable lengths
NOTE
Cable lengths longer than 60 m require a heavier gauge wire if the
power supply drops below 11 V.