Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
6 - 6
Manufacturing settings
When the tests of
Table 6.1 have been completed, the pump can be returned to the manufacturing
settings (if required) as shown in sequence in
Table 6.2.
The settings marked with an asterisk (
*
) in
Table 6.2 are intermediate settings that are initially
selected in order to allow several other associated settings to appear in the Configuration
mode. Ultimately, the final sequence of Manufacturing settings are selected as shown in the
latter part of
Table 6.2.
Manufacturing
settings
Option
Setting
Configuration mode:
1
Syringe type
BD PLASTIPAK
2
Syringe size
NOT LOCKED
3
Lock program values
NO
4
Max infusion rate
200 ml/h
5
Select pump modes
ALL MODES AVAILABLE *
6
Preset volume mode with time
YES
7
KVO Rate
0.5 ml/h
8
Allow mass units option
YES *
9
Infusion units mg/kg/h
YES
10
Infusion units mg/kg/min
YES
11
Infusion units µg/kg/h
YES
12
Infusion units µg/kg/min
YES
13
Infusion units mg/h
YES
14
Infusion units µg/h
YES
15
Show rate in ml/h while infusing
YES
16
Use pressure transducer
YES
17
Display pressure bargraph
YES
18
Allow bolus while running
NO
19
Allow rate change while running
NO
20
Intermittent mode start delay
YES
21
RS232 baud rate
9600
22
Communication mode
NONE
23
Pump ID
GRASEBY_3200 (see Note over page)
24
Set key beep volume
5
25
Pressure units
mmHg
Setup mode:
1
Infusion mode
INTERMITTENT
*
2
Dose volume
0.2 ml
3
Dose duration
STAT
4
Dose cycle time
4 hours 0 minutes
5
Background rate
0.00 ml/h
6
Start delay
0 hours 0 minutes
7
Pressure limit
75 mmHg
8
Infusion mode
PRESET VOLUME
*
9
Infusion rate
1.0 ml/h
10
Preset volume
0.2 ml/h
11
Infusion mode
CONTINUOUS
12
Infusion units
mg/kg/h
*
13
Patient body weight
0.40 kg
14
Drug mass
1.0 mg
15
Drug volume
60 ml
16
Infusion rate
0.01 mg/kg/h
17
Infusion units
ml/h
(contd.)
Table 6.2 Manufacturing settings
Содержание Graseby 3200
Страница 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Страница 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Страница 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 70: ...CHAPTER 5 FAULT CODES CLEANING RENEWAL of FUSES and REPAIRS 3200 In line Pressure Syringe Pump ...
Страница 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Страница 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Страница 95: ...Smiths Medical International Ltd 7 4 Issue 5 August 2004 3200 Service Manual 3200 Illustrated Parts List ...
Страница 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Страница 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Страница 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...