Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
1 — 4
100 V to 240 V at 50/60 Hertz, 40 VA.
The power supply uses primary switching in order to utilise the AC supplies of most
countries.
Sealed lead acid, rechargeable (Cyclon, 3 off). Smiths Medical recommend that the
batteries are checked at least annually (see page 5-10).
More than 2.5 hours of normal pump operation when the batteries are fully charged.
With the AC supply connected, up to 14 hours are required to fully recharge low voltage
batteries.
3200 DC variant pump.
10 V to 28 V DC at a maximum of 4 amps input supply (see Chapter 9).
BD Plastipak...
5, 10, 20, 30/50 and 50/60 ml.
Injectomat...
10 ml and 50 ml.
IMS Pumpjet...
30 ml, prefilled.
Monoject...
5, 10, 20, 30/35 and 50/60 ml.
Braun Omnifix...
5, 10, 20, 30 and 50/60 ml.
Terumo...
5, 10, 20, 30/35 and 50/60 ml.
BD Plastipak (A)... 5, 10, 20, 30/35 and 50/60 ml.
The BD Plastipak (A) syringe accommodates the American manufactured 10 ml syringe,
which differs slightly in length to the European version. All other syringe sizes in the BD
Plastipak selection are identical to the standard BD Plastipak syringes.
Braun Perfusor...
Optional (see Chapter 8).
AC power
supply:
Battery type:
Specification
Rates
Increments
0.05 to 1.0 ml/h...
0.01 ml/h
1 to 200 ml/h...
0.1 ml/h
0.01 to 999 mg/kg/h
0.01 to 999 mg/kg/min
0.01 to 999 µg/kg/h
0.01 to 999 µg/kg/min
0.01 to 999 mg/h
0.01 to 999 µg/h.
Patient weight...
0.4 to 200 kg.
Drug mass...
1 µg to 999 mg.
Drug volume...
1 ml to 60 ml.
0 to 999.9 ml in 0.1 ml increments.
Infusion rates
and increments:
Mass unit
programming
range:
Volume infused
counter:
Specification
Battery life:
Syringe brands
and sizes:
DC input
operation:
Содержание Graseby 3200
Страница 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Страница 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Страница 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 70: ...CHAPTER 5 FAULT CODES CLEANING RENEWAL of FUSES and REPAIRS 3200 In line Pressure Syringe Pump ...
Страница 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Страница 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Страница 95: ...Smiths Medical International Ltd 7 4 Issue 5 August 2004 3200 Service Manual 3200 Illustrated Parts List ...
Страница 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Страница 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Страница 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...