Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual
8 — 1
Braun Perfusor conversion
CHAPTER 8
BRAUN PERFUSOR CONVERSION
The 3200 can, if required, easily be converted to use the ‘Braun Perfusor 50 ml syringe’.
The Perfusor conversion kit part number is 0131-0048.
The conversion procedures are detailed below (see
Figure 8.1).
Open the pump case (see page 2-11) and temporarily displace the Distribution board
assembly (see
Figure 7.1) by removing the screw at the bottom of the assembly.
Remove the standard flag
from the top of the half nut (retain the standard flag for possible
future use). Fit the elongated Perfusor 50 ml nearly empty flag (see page 7-25) by inserting
the flag into the grooves on the half nut (see
Figure 8.1). Ensure that the flag is fitted so
that the small amount of flag bias is in the direction of the arrow shown in
Figure 8.1.
Note:
If not already biassed, bend/bias the end of the flag in the direction of the arrow
shown in
Figure 8.1 a distance of approximately 2 mm, this will ensure that the flag
(when it moves fully to the left) does not touch the left hand post.
Carefully lift the left end of the inner metal tube and fit the nylon spacer tube (part number
0131-0051) over the tube. The spacer tube acts as an end of travel stop when the half nut
assembly is moved to the left.
Fit the self adhesive half-round plunger clamp plate (part number 0131-0073) to the left
hand side of the plunger clamp. It must be fitted to the vertical face of the plunger clamp
so that it covers both quadrant vanes (see
Figure 8.1).
Reassemble the case (see page 2-11).
The Braun Perfusor 50 ml syringe is selected for use (on the pump) by using a special
configuration command.
First select the configuration mode (see page 2-1) and then select and display the syringe
brand screen.
Simultaneously press the following three buttons:
ALARM, PURGE
and
OFF.
The pump will then switch to the Braun Perfusor syringe mode and other syringe brands
will automatically be disabled. The pump syringe display will show:
BRAUN PERFUSOR 50 ml
and the
CHANGE
button, whilst in this mode is ineffective.
Fit the self adhesive ‘P’ label (part number 0131-0065) to the front of the pump in the area
just above and to the right of the product name (see
Figure 8.1). This label acts as a visual
reminder that the pump must only be used with the Braun Perfusor syringe.
Syringe conversion procedures
Introduction
Nearly empty
flag
conversion
Spacer tube
fitment
Plunger clamp
fitment plate
Braun
Perfusor
selection
‘P’ label
fitment
Содержание Graseby 3200
Страница 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Страница 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Страница 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 70: ...CHAPTER 5 FAULT CODES CLEANING RENEWAL of FUSES and REPAIRS 3200 In line Pressure Syringe Pump ...
Страница 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Страница 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Страница 95: ...Smiths Medical International Ltd 7 4 Issue 5 August 2004 3200 Service Manual 3200 Illustrated Parts List ...
Страница 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Страница 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Страница 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...