Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
2 — 14
If the occlusion thrust requires adjustment, the following procedures must be completed:
1.
Switch the pump off and disconnect the AC supply.
2.
Take the casing apart (see page 2-11).
3.
If necessary, rotate the leadscrew to reveal the grub screw located on the occlusion
adjusting nut. Loosen the grub screw with a 1.5 mm hexagonal key.
4.
Alter the setting of the occlusion adjusting nut as necessary. One full turn of the
adjusting nut using a 50\60 ml syringe gives approximately 2.73 kg (369 mmHg) of
adjustment.
Rotating the adjusting nut to decrease the tension on the leadscrew spring. This decreases the
pump's occlusion setting.
Rotating the adjusting nut to increase the tension on the leadscrew spring increases the
pump's occlusion setting.
5.
Tighten the grub screw to a torque of 15 ± 2cNm.
6.
Temporarily assemble the two halves of the pump, being careful not to trap any
leads.
7.
Complete a thrust check (see page 2-13).
8.
To obtain the thrust required it may be necessary to repeat steps (2) to (7) above.
9.
Assemble of the pump casing (see page 2-11).
10.
Complete tests No. 9 and 10 detailed in the functional test procedures (see page 6-3).
Occlusion thrust
Thrust
adjustments
Содержание Graseby 3200
Страница 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Страница 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Страница 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 70: ...CHAPTER 5 FAULT CODES CLEANING RENEWAL of FUSES and REPAIRS 3200 In line Pressure Syringe Pump ...
Страница 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Страница 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Страница 95: ...Smiths Medical International Ltd 7 4 Issue 5 August 2004 3200 Service Manual 3200 Illustrated Parts List ...
Страница 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Страница 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Страница 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...