6 - 2
5(vii)
Check that the change rate facility The rate of infusion changes.
functions as described in the
Instruction Manual.
5(viii)
Whilst infusing, press
STOP
.
The pump will enter its set-up mode.
5(ix)
Press
START
to commence the
The alarm sounds intermittently,
infusion and then move the plunger
the
ALARM
LED flashes and...
clamp to its open position.
PLUNGER CLAMP OPEN
is displayed.
5(x)
In the above alarm state [5 (ix] press
The alarm is silenced and the pump
the
ALARM
key.
goes into its set-up mode.
6(i)
AC failure.
Press
START
to start the infusion
The alarm will sound intermittently
and then switch off or remove the
and ...
AC POWER REMOVED
AC supply externally.
will be displayed, but the pump
will continue running under battery
power.
6(ii)
Reconnect the AC supply.
The
AC
LED lights.
Functional tests
Table 6.1 Functional tests
(contd.)
Step
Test
Method
Correct result
5(i)
Keyboard tests.
Press
OFF
.
The display is blank but the
AC
AC connected.
LED remains lit.
5(ii)
Press
ON
.
As above (3), Initial power on.
5(iii)
Check that the pump can be
Confirm display messages shown
programmed by using the
▲ ▼
in the Instruction Manual, and also
and
ENTER
keys as described
confirm that the keys change
in the Instruction Manual.
the settings as required.
5(iv)
Check that the
History-
display/reset
Display shows volume infused
keys operate correctly as described
since last reset, then resets to zero
in the Instruction manual.
when the
reset
key is pressed.
5(v)
Fit a syringe and move the plunger
The syringe type and size is
clamp to the closed position. Prog-
displayed. The
START
LED flashes
ramme the pump, then press
START
. and the display running indicator
arrows flash.
5(vi)
Check that the bolus facility functions
ALARM
and
START
LEDs flash
as described in the Instruction
and the bolus delivered is displayed.
Manual.
(contd.)
Содержание Graseby 3200
Страница 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Страница 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Страница 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 70: ...CHAPTER 5 FAULT CODES CLEANING RENEWAL of FUSES and REPAIRS 3200 In line Pressure Syringe Pump ...
Страница 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Страница 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Страница 95: ...Smiths Medical International Ltd 7 4 Issue 5 August 2004 3200 Service Manual 3200 Illustrated Parts List ...
Страница 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Страница 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Страница 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...