Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual
2 — 11
Opening the pump casing
The casing must be opened to carry out adjustments that may be required to the occlusion
thrust and for various repair procedures.
Repair procedures are detailed in Chapter 5.
Each time the casing separated and reassembled, the syringe size functional tests detailed
on page 6-1 must be completed.
1.
Disconnect the AC power connector and utilising a scratch free flat surface, turn the
pump over to gain access to the base.
2.
Undo and retain the six screws that hold the casing halves casing together. One of
the screws is situated in a channel in the rear cover.
3.
From the top of the pump carefully ease the casing halves apart, taking care not to
put any strain on the internal connecting cable looms that form a hinge between the
casing halves.
Being careful not
to trap any leads, assemble the casing by reversing steps (2) and (3) detailed
above. Ensure that the case halves have ‘snapped’ together and that the front and rear mating
edges are equal and parallel. Tighten the screws to a torque of between 70 and 75 cNm in the
order shown in the
Figure 2.1.
Taking the
casing apart
Disassembly and assembly of casing
WARNING
The following procedures must only be carried out by qualified technicians.
ELECTRIC SHOCK HAZARD
The pump
must
be disconnected from the AC power supply before opening
the casing.
Assembly
1
3
2
5
GM0595-B
4
6
Figure 2.1 Case fixing screw tightening order
Содержание Graseby 3200
Страница 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Страница 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Страница 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 70: ...CHAPTER 5 FAULT CODES CLEANING RENEWAL of FUSES and REPAIRS 3200 In line Pressure Syringe Pump ...
Страница 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Страница 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Страница 95: ...Smiths Medical International Ltd 7 4 Issue 5 August 2004 3200 Service Manual 3200 Illustrated Parts List ...
Страница 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Страница 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Страница 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...