Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
Appendix, page A - 4
SSF, Appendix
Setting the size sensor flag
1.
Using the 3200 size sensor test gauges (see page A-5), fit the white 20 ml minimum
sensor gauge (
part number 0131-0170) into pump's cradle and observe whether (a)
or (b) occurs:
a.
If the pump indicates that a 20 ml gauge is fitted, complete steps (2) to (5).
b.
If no indications are given, complete steps (6) to (8).
2.
Depending upon the type of pump being set-up, if either the
20ml
LED is
illuminated, or the display shows
20
, rotate the grub screw counter-clockwise until
the corresponding LED/ display is extinguished or disappears.
3.
Turn the grub screw slowly clockwise until the LED/ display illuminates or appears,
indicating that a 20 ml syringe is fitted.
4.
At position X (see
Figure A.3 Size Sensor Flag, general details), alternately press
and release the SSF several times and ensure that when the SSF returns to it's rest
position, it
always
displays that a 20 ml syringe is present.
Note
: It is acceptable that while the SSF is under force, the size of the syringe displayed
alters. However, it must always return to the original 20 ml size displayed when the force
is released.
5.
If the pump does toggle between different syringe sizes, turn the grub screw
very
slightly clockwise and complete step (4) again until the 20ml LED/ display is stable.
Procedures if (b) above occurs
6.
Depending upon the type of pump being set-up, if the
20 ml
LED is
not
illuminated,
or the display does
not
show
20
, rotate the grub screw slowly clockwise until the
LED/ display illuminates or appears indicating that a 20 ml syringe is fitted.
7.
At position X (see
Figure A.3 Size Sensor Flag, general details), alternately press
and release the SSF several times and ensure that when the SSF returns to it's rest
position, it
always
displays that a 20 ml syringe is present.
Note
: It is acceptable that while the SSF is under force, the size of the syringe displayed
alters. However, it must always return to the original 20ml size displayed when the force is
released.
8.
If the pump does toggle between different syringe sizes, turn the grub screw
very
slightly clockwise and repeat step (7) until the 20 ml LED/ display is stable.
Apply a small amount of Loctite 414 to the anti-tampering grub screw cap and fit it.
When the grub screw in the SSF has been set, the remaining procedures required to
complete the size sensor testing may be completed. Press and release the SSF
each time a new gauge is placed in the pump's cradle.
Содержание Graseby 3200
Страница 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Страница 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Страница 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 70: ...CHAPTER 5 FAULT CODES CLEANING RENEWAL of FUSES and REPAIRS 3200 In line Pressure Syringe Pump ...
Страница 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Страница 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Страница 95: ...Smiths Medical International Ltd 7 4 Issue 5 August 2004 3200 Service Manual 3200 Illustrated Parts List ...
Страница 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Страница 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Страница 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...