Smiths Medical International Ltd.
3200 Service Manual
Issue 5 (August 2004)
8 — 2
Braun Perfusor conversion
To reselect the various syringe brands carry out the following mechanical and
programming procedures:
1.
Take the case apart (see page 2-5).
2.
Replace the original standard nearly empty flag.
3.
Remove the nylon spacer from the left hand side of the inner metal tube.
4.
Remove the adhesive plunger clamp plate.
5.
Reassemble the case.
6.
Remove the adhesive ‘P’ label from the front of the pump.
Reselecting ‘various’ syringe brands
Mechanical
procedures
Programming
procedures
1.
With the pump switched
ON
and
BRAUN PERFUSOR 50 ml
displayed within the configuration mode (see page 2-1), simultaneously press the
following three buttons:
ALARM, PURGE and OFF
in order to set the configuration mode so that the various brands of syringe may be
selected.
2.
Set the pump to the brand of syringe that is going to be used and then press the
STOP
button in preparation for the next infusion.
Nylon
spacer
tube
Nearly
empty
flag
Adhesive
label
Adhesive
plunger
clamp plate
GM0420-B
Flag
bias
5ml
20ml
50/60ml
Graseby
Syringe pump
3200
P
OFF
display
reset
STOP
10ml
30ml
History
Figure 8.1 Braun Perfusor conversion: parts required
Содержание Graseby 3200
Страница 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Страница 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Страница 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 70: ...CHAPTER 5 FAULT CODES CLEANING RENEWAL of FUSES and REPAIRS 3200 In line Pressure Syringe Pump ...
Страница 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Страница 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Страница 95: ...Smiths Medical International Ltd 7 4 Issue 5 August 2004 3200 Service Manual 3200 Illustrated Parts List ...
Страница 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Страница 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Страница 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...