Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual
3 — 5
Software
When the pump is switched on with a disposable syringe extension set fitted, the user is
alerted that an auto zero has not been carried out.
An alarm is generated if an extension set calibration is unsuccessful.
A range of 0 to 1000 mmHg is specified for the extension set, with an offset of +/-100
mmHg. If this offset is exceeded, an unsuccessful calibration alarm is generated.
If the extension set is removed from its housing while the pump is running an alarm condi-
tion will be activated.
When under the control of an external computer, an alarm is generated if a computer failure
or disconnection is detected.
Cannot zero
extension set
Cannot
calibrate
extension set
Communications
(RS232) failure
The 3200 self tests include the following:
•
ROM test (CRC-16),
•
RAM test,
•
power supply voltage test,
•
keyboard test (this test checks for shorted keys),
•
stack usage test, and
•
motor windings continuity test.
The program is held in a 256 Kbyte EPROM.
The configuration data and settings are stored in the processor’s EEPROM.
A 32 Kbyte RAM chip holds the history. The history will be lost if the AC supply to the pump
is switched off or disconnected and the internal batteries unplugged.
Self tests
Design
Software
Содержание Graseby 3200
Страница 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Страница 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Страница 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 70: ...CHAPTER 5 FAULT CODES CLEANING RENEWAL of FUSES and REPAIRS 3200 In line Pressure Syringe Pump ...
Страница 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Страница 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Страница 95: ...Smiths Medical International Ltd 7 4 Issue 5 August 2004 3200 Service Manual 3200 Illustrated Parts List ...
Страница 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Страница 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Страница 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...