Smiths Medical International Ltd.
Issue 5 (August 2004)
3200 Service Manual
1 — 5
Specification
In-line occlusion
pressure range:
Internally adjustable
occlusion pressure:
Factory set occlusion:
Display languages
available:
Dimensions:
Weight:
Temperature
range:
Drive accuracy:
History:
Electrical safety:
Design standards:
UK patent number:
Printer protocol:
Using a 30 ml sized syringe:
0 to 7.42 kg (1000 mmHg).
0 to 1355 cmH
2
O.
0 to 133.3 kPa.
0 to 1316 mBar.
0 to 19.34 psi.
Using a 50 ml syringe, 0 to 4.83 kg (0 to 650 mmHg).
Using a 50 ml syringe, 1.85 kg (250 mmHg) to 4.83 kg (650 mmHg).
3.5 kg (471 mmHg) to 4.2 kg (565 mmHg).
Dutch.
English.
French.
German.
Italian.
Norwegian.
Spanish.
350 x 195 x 115 mm, with pole clamp fitted and plunger clamp closed.
Not exceeding 3.5 kg including the batteries and pole clamp.
Operating conditions:
+5°C to +40°C, 30 to 75% Rh, 700 to 1060 hPa.
Storage conditions:
-40°C to +70°C, 30 to 90% Rh, 700 to 1060 hPa.
±2%.
1500 events can be stored.
Class II
Type CF
Drip proof IPX1.
BS5724 Part 1
IEC601 Part 1
VFG1046/1984.
2229497.
Serial, 8 bits, no parity, 1 stop bit.
Specification
(contd.)
Содержание Graseby 3200
Страница 15: ...CHAPTER 1 INTRODUCTION FEATURES and SPECIFICATION 3200 In line Pressure Syringe Pump ...
Страница 22: ...CHAPTER 2 CONFIGURATION DIAGNOSTICS and OCCLUSION THRUST 3200 In line Pressure Syringe Pump ...
Страница 37: ...CHAPTER 3 FUNCTIONAL DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 43: ...CHAPTER 4 CIRCUIT DESCRIPTIONS 3200 In line Pressure Syringe Pump ...
Страница 70: ...CHAPTER 5 FAULT CODES CLEANING RENEWAL of FUSES and REPAIRS 3200 In line Pressure Syringe Pump ...
Страница 83: ...CHAPTER 6 FUNCTIONAL TESTS and MANUFACTURNG SETTINGS 3200 In line Pressure Syringe Pump ...
Страница 91: ...CHAPTER 7 ILLUSTRATED PARTS LIST 3200 In line Pressure Syringe Pump ...
Страница 95: ...Smiths Medical International Ltd 7 4 Issue 5 August 2004 3200 Service Manual 3200 Illustrated Parts List ...
Страница 105: ...CHAPTER 8 BRAUN PERFUSOR CONVERSION 3200 In line Pressure Syringe Pump ...
Страница 108: ...CHAPTER 9 DC INPUT VERSION of 3200 3200 In line Pressure Syringe Pump ...
Страница 111: ...APPENDIX FITTING a MODIFIED SIZE SENSOR FLAG 3200 In line Pressure Syringe Pump ...