S E C T I O N 5
•
O u t o f t h e B o x - A s s e m b l y
Level 1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
11
Step 5 - To Set Up the 3-Liter (3L) Pressure Accessory
1
Disconnect the black pneumatic tubing from each In-Line Pneumatic
Fitting (
b
). Press on the red ring closer to the 1L pressure chamber,
and gently pull on the black tubing.
2
Remove the black pneumatic tubing from the J-clips (
c
) on the sides
of the pole.
3
Make sure the Adapter Assembly’s valve handle is in the OFF/
VENT position (
d
).
4
Insert the black tube from the Adapter Assembly into the In-Line
Pneumatic Fitting (
b
).
5
Connect the Adapter Assembly (
g
) to the 3L Pressure Cuff (
f
) by
connecting the male and female Luer connectors (
e
) securely.
6
Slide the 3L Fluid Bag into the Pressure Cuff, and hang them from
the Fluid Bag Hanger.
7
Repeat steps 3-6 if you are using a second 3L Pressure Cuff.
Step 6 - Disinfect the Recirculating Solution Reservoir
1
Remove the fill-port plug (
a
) on the reservoir.
2
Prepare a 0.3% hydrogen peroxide/sterile water solution for the
reservoir. Mix 140 ml of 3% hydrogen peroxide solution and 1,260 ml
of sterile water.
3
Fill the reservoir with 1.4 liters of 0.3% hydrogen peroxide/ sterile
water solution.
4
Replace the fill-port plug.
5
Insert a Disposable Set into the Fluid Warmer.
6
Insert the power cord into a properly grounded receptacle.
7
Turn the Fluid Warmer ON. Let the solution circulate for a
30-minute disinfection period.
8
Turn the Fluid Warmer OFF.
9
Empty the reservoir.
10
Remove the Disposable Set and discard according to established
hospital procedures.