S E C T I O N 8
•
O p e r a t i o n
24
Level
1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
Check Disposables Mode
The Check Disposables mode of the Fluid Warmer indicates a missing or
improperly installed Disposable Set.
Mode characteristics
• The Check Disposables yellow LED (
a
) on the Fluid Warmer's
Display Panel illuminates.
• An audible attention signal beeps.
• Reservoir solution circulation is stopped; fluid warming stops.
• Pressure Chambers continue to operate.
To clear this mode
Install a Disposable Set or check the Disposable Set installation as
follows:
• Check Heat Exchanger for proper installation in Block
1
, and in
Interlocks
2
and
4
.
• Press Heat Exchanger down firmly in Block
1
to secure in
O-Ring.
• Press Interlock
2
tab down firmly to engage the Interlock switch.
• Check Gas Vent installation in Interlock
3
.
Add Recirculating Solution Mode
The Add Recirculating Solution mode of the Fluid Warmer indicates that
the solution level in the recirculating solution reservoir is below its
minimum level.
Mode characteristics
• The yellow Add Recirculating Solution LED (
b
), on the Fluid
Warmer's Display Panel, illuminates.
• An audible warning signal beeps.
• Solution circulation stops; fluid warming stops.
• Pressure Chambers continue to operate.
• Fluid flow to the patient continues.
To clear this mode
• Add recirculating solution to the Fluid Warmer's reservoir.
a
3
3
b