S E C T I O N 1 5
•
S p e c i f i c a t i o n s
50
Level
1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
SECTION 15
Specifications
System Specifications
Standard Compliance
Guidelines
Product Safety
EN 60601-1, UL 2601-1
EMC
EN 60601-1-2, FCC 47 CFR Part
15, Class B
Enclosure Protection
IEC 60529 IP Code: IPX1
Fluid Warmers
ASTM F2172-02
Physical
Dimensions
Height, Overall
1.83 m
(6 ft)
Width, Overall
48.3 cm
(19 inches)
Depth, Overall
48.3 cm
(19 inches)
Weight Assembled, Dry
38.1 kg
(84 pounds)
Recirculating Solution Capacity
1.4 L
(0.37 gallons)
Air Source Pressure
0-500 mmHg
Environmental
Temperature
Humidity [%]
Operation
10
o
C to 45
o
C
30 to 75
Transportation
0
o
C to 70
o
C
30 to 75
Storage
0
o
C to 70
o
C
30 to 75
Thermal
Temperature
Temperature Set Point
41.7
o
C ± 0.3
o
C
Over Temperature Set Point
43.9
o
C ± 0.1
o
C
Electrical
Type
Protection Against Electrical Shock
Class 1 Equipment, Type BF
Mode of Operation
Continuous
Type of Current
Alternating
Ingress Protection Rating
IPX1
MAINS Power Input
115VAC, 50/60 H
Z
, 12.0 Amps
MAINS Auxiliary Power Supply Output
115VAC, 50/60 H
Z
, 1.5 Amps