S E C T I O N 9
•
O p e r a t i n g I n s t r u c t i o n s
34
Level
1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
D–Change the Fluid Bag
1
Close the clamp under empty bag.
2
Push the Pressure Chamber Toggle Switch to the minus (
-
)
unpressurized position to release pressure in the pressure
chambers.
3
Open the pressure chamber door and remove the fluid bag from
the fluid bag hanger.
4
Remove the spike from the empty fluid bag.
5
Hang a new fluid bag.
6
Spike the fluid bag. Be sure tubing is not twisted or kinked.
7
Close and latch Pressure Chamber doors.
8
Push the Pressure Chamber Toggle Switch to the plus (+)
pressurized position
9.3 After Use
1
Close all clamps.
2
Push the Pressure Chamber Toggle Switch to the minus (
-
)
unpressurized position to release pressure in the pressure chambers.
3
Turn the Fluid Warmer OFF.
4
Open the Pressure Chamber door and remove the fluid bag.
5
Properly discard the Disposable Set in containers marked for
biohazardous materials. Dispose by incineration, or follow hospital
policies and procedures applicable for the disposal of biohazardous
material.
6
Visually check the condition of the device. Remove from service any
unit that shows physical damage.
7
Clean the device with warm soapy water.