S E C T I O N 1 1
•
T e s t i n g
38
Level
1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
3
Gas Vent Interlock - Pull the top of the Gas Vent Assembly from
the #3 Block. The Check Disposables indicator illuminates and
the audible alarm beeps.
Over Temperature Test
Do the following:
1
Insure that the Fluid Warmer is at operating temperature (41°C).
2
Press and hold the Over Temperature Test button (
a
).
3
The Over Temperature LEDs illuminate, and an audible alarm
beeps.
4
Release the Over Temperature Test button.
5
Over Temperature LED and audible alarm signal remains active.
Clear the Alarm mode
• Turn the Fluid Warmer OFF and then back ON.
Fluid Warmer Alarm Signal Test
The Alarm Test button is used to confirm proper operation of the visual
and audible alarm indicators.
Do the following
1
Press and hold the Alarm Test button (
b
).
2
The Check Disposables, Add Recirculating Solution, and Over
Temperature LEDs illuminate and an audible alarm beeps.
3
Release the Alarm Test button.
4
Over Temperature LED remains lit, and the audible alarm continues
to beep.
Clear the Alarm mode
• Turn the Fluid Warmer OFF and then back ON.
a
b