S E C T I O N 7
•
C o n t r o l s a n d D i s p l a y s
16
Level 1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
Fluid Warmer Power and Alarm Test Panel
The Power and Alarm Test Panel is located on the front of the Fluid
Warmer directly above the reservoir fill-cap. This panel contains four
pressure-sensitive buttons that are activated when pressed. Refer to the
Power Alarm Test Panel figure, whose numbered call-outs correspond
to descriptions of the buttons and the functions they perform.
Button/Function
1 Power ON Button
The green button on the top left of the Power and Alarm Test
Panel powers on the device. The green Automatic Operation LED
on the Fluid Warmer Display Panel illuminates when the Power
ON button is activated, and a
Disposable Set is in place.
2 Power OFF Button
The Power OFF is the orange button to the right of the Power ON
button on the Power and Alarm Test Panel. This button turns off
power to the unit. The green Automatic Operation LED on the
Display Panel will turn off when this button is pressed.
3 Over Temperature Test Button
The Over Temperature Test is used to confirm the proper
operation of the Over Temperature circuitry. Testing the circuitry
requires that the Fluid Warmer is at operating temperature
(41°C). Once this is established, press and hold the button until
the alarm activates. Then, release the button. The Over
Temperature alarm continues to function. Clear the alarm mode
by turning the device off, then back on.
4 Fluid Warmer Alarm Signal Test Button
The Fluid Warmer Alarm Signal Test is used to confirm proper
operation of the visual and audible alarm indicators. Press and
hold this button to test circuitry. Then, release the button. The
Over Temperature alarm continues to function. Clear the alarm
mode by turning the device off, then back on.
5 Reservoir Capacity
Capacity for recirculating solution reservoir is 1.4 liters.
Use recirculating solution. Do not exceed maximum capacity.