S E C T I O N 9
•
O p e r a t i n g I n s t r u c t i o n s
Level 1
®
NORMOFLO
®
H-1129 Irrigation Fluid Warmer | Operator’s Manual
33
6
Turn the valve handle on the Adapter Assembly to the ON position
(
c
). The Pressure Cuff begins to inflate. The Pressure Gauge drops to
zero temporarily, and then steadily increases to the desired pressure.
The unit is ready for use.
7
To deflate the Pressure Cuff, turn the valve handle (
c
) back to the
OFF/VENT position (b).
C– Adjust the Automatic Pole
WARNING
• Do not overextend Automatic Pole; if Automatic Pole is
overextended the Fluid Warmer may become unstable.
CAUTION
• Avoid contact with surgical lights or other overhead fixtures.
Maximum Transport Load: 12 liters at maximum height of 2.25
meters (7 ft 4 inch), 6 liters at maximum height of 2.54 meters
(8 ft 4 inch) from floor to bag hanger.
• To raise the pole, push and hold the Membrane Switch Up Arrow
located on the left side of the Automatic Pole until the pole reaches
the desired height, or until the pole reaches its highest position;
release the Membrane Switch.
Note: Observe that the Disposable Set is not snagged as the Pole is
extended.
• To lower the pole, push and hold the Membrane Switch Down
Arrow until the pole reaches the desired height, or until the pole
reaches it lowest position; release the Membrane Switch.
c
b
g
2.25M / 12 LITERS
2.54M / 6 LITERS